Doppler Lectures Flashcards

1
Q

Presentation of flow information in color, superimposed on two dimensional grey scale image.

A

COLOR DOPPLER

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Color doppler is formed of:

A

PULSED DOPPLER

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Number of ultrasound pulses per line of color.

A

PACKET SIZE

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q
  • Better velocity measurements.
  • Better color sensitivity.
  • Worse temporal resolution (decreased frame rate)
A

LARGER PACKET SIZES

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

INFORMATION YIELDED:

A
  • Pressure of flow.
  • Flow direction.
  • Reflector velocity. (qualitive)
  • Reflector variance.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

ADVANTAGES:

A
  • Provides two dimensional information.
  • Facilitates examination
  • Provides info. - direction, character, relative velocities, amplitude of flow
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

DISADVANTAGES:

A
  • Poorer temporal resolution.
  • Subject to aliasing.
  • Cannot measure exact velocities.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

COLOR ALIASING DUE TO HIGH VELOCITIES OR COLOR SCALE SETTING TOO LOW.

A
  • Colors wrap around color bar.

- Shows positive as negative.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Methods of displaying colors in relation to velocity.

A

COLOR ASSIGNMENT MAPS

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q
  • Darker, duller color represent slower velocities .

- Brighter, lighter colors represent faster velocities,

A

VELOCITY MAP

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q
  • Another color is added (usually green)
  • With increase in variance, brightness of green increases.
  • Color green signifies disturbed, turbulent flow-Used in cardiac imaging.
A

VARIANCE MAP

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

COLOR COMPONENTS

A

HUE- COLOR
SATURATION - DILUTED W/WHITE
LUMINANCE - BRIGHTNESS OF HUE AND SATURATION

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

AUTOCORRELATION

PHASE SHIFT AUTOCORRELATION

A
  • Used for color doppler.
  • Calculates AVERAGE shift of echos from same packet.
  • Similar, but faster than FFT.
  • QUALITATIVE INFO.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q
  • Newer method of Doppler shift analysis.
  • Faster than FFT.
  • Can pinpoint specific echo out of many in time.
A

TIME DOMAIN PROCESSING

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

COLOR POWER DOPPLER

COLOR DOPPLER ENERGY, ULTRASOUND ANGIO, COLOR POWER ANGIO

A
  • Estimates power of doppler shift in specific region.
  • Power of shift relates to concentration of reflector.
  • Info based on strength or power of reflectors.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

ADVANTAGES OF POWER DOPPLER

A
  • Not angle dependent.
  • Not subject to aliasing.
  • Greater sensitivity to flow.
17
Q

DISADVANTAGES OF POWER DOPPLER

A
  • Not direction sensitive.
  • Cannot detect velocity.
  • Cannot depict flow characteristics.
  • Takes more time.