Chapter 36 Flashcards
Leakage radiation must be less then ____mR/hr at a distance of 1 m from the protective housing
100
The SID indicator must be accurate within __% of the indicated SID.
2
The x-ray beam and the light must beam must coincide within ___% of the SID
2
The PBL must be accurate to within ___ of the SID
2
The variation in x-ray intensity should not exceed ___%
5
The maximum acceptable variation in linearity is ___% from one mA station to an adjacent mA station.
10
The SSD must not be less than ___cm on stationary fluroscopes and not less then ___ cm on mobile fluroscope.
38, 30
___ time produces an audible signal when the fluroscopc time has exceeded 5 minutes
cumulative
What device is used to cover the opening on the side of a fluroscopic table when the bucky tray is moved out the way?
bucky slot cover
A ___ of at least 0.25mm Pb equivalent should be positioned between the fluroscopist and the patient
protective curtain
____ radiation is the useful beam.
primary
What are the two types of secondary radiation:
scatter and leakage
The intensity of scatter radiation 1 m from the patient is approximately __% of the intensity of the useful beam at the patient.
0.1
When PBL is activated, the collimators are automatically adjusted so that the radiation field matches the ____.
film size
What are the two type of filtration?
inherent and added
The glass envelope encasing the
x-ray tube, the insulating oil, and the glass window are examples of what type of filtration?
inherent
Inherent materials amounts to approximately __mm of aluminum equivalent.
0.5