Automated Hematology Instrumentation Flashcards
Before running the sample
-CBC specimens must be checked for clots (visually, with applicator
sticks, or by automated flags), significant hemolysis, and lipemia before
reporting results
-Processing either automated or manual, should be done within 8 hours
but never longer than 24 hours after sample collection
-Samples must be thoroughly mixed before testing
Automated Blood Cell-
Counting Instruments
1.) electrical Impedance
◦ Analyzes the resistance created by each cell passing through an electrical
field
◦ Coulter principle – increased resistance occurs when poorly conductive
blood cells pass through aperture.
◦ Ex: Abbott Diagnostics, Beckman Coulter Inc., and Sysmex Corporat
Automated Blood Cell-
Counting Instruments
2.) optical light scatter
◦ Analyzes the scatter of light (forward and side) detected by each cell
passing through a beam of light (optical or laser)
◦ Ex: Siemens Healthcare, Sysmex Corporation, and Abbott Diagnosti
Electrical impendane Instruments
1st aliquot
RBC/plt dilution chamber
* External electrode
* 3 apertures (ea w/ internal electrode)
3 RBC counts are obtained, compared, and evaluated. If agreement, the
reported RBC count is an average of the 3 counts.
If RBCs are larger then normal then
Shift to the right
If RBCs are smaller then normal then
Shift to the left
Histogram
Size distribution curve
PLT size
Less then 20um
RBC size
less then 36 Um
Blood aspirated in
2-4 aliquots
1st aliquot determinations
Data determines:
◦ RBC count
◦ MCV
◦ RDW-CV and RDW-SD
◦ Plt count
◦ MPV
Calculated parameters:
◦ Hct
◦ MCH
◦ MCHC
Gaussian Curve Abnormal Curve
2nd Aliquot determinations
Delivered to what chamber
Lytic agents lyses what and converts what
Delivered to WBC/hemoglobin dilution chamber
◦ Lytic agent lyses RBCs and converts released hgb to cyanmethomoglobin,
and shrinks the leukocyte cell membrane and cytoplasm… allowing WBC
count to represent cell volume rather than cell size
2nd aliquot
WBC count
Hgb count
◦ WBC count measured by electrical impedance from 3 apertures and
reported count represents the average.
◦ Hgb determined by absorbance reading at 525 nm (Beer’s La
Lymphocytes, Monocytes basophils and eosinophils, neutrophils curve
Lymphocytes on left
Mono, eosino, baso in middle
Neutrophils on right
3rd aliquot
steps 1,2, and 3
delivered to the orbital mixing chamber
1. Blood mixed with heated lysing agent to remove RBCs
2. Stabilizing agent added to preserve WBCs
3. Cells sent through the volume-conductivity-scatter (VCS) flow cell by
hydrodynamic focusing for 5-part differenti
3rd aliquot important cell measurements
- Cell volume – by impedance
- Cell conductivity – by electromagnetic probe. Determines physical and chemical components
- Cell’s light scatter characteristics - determines internal contents and cell surface and size
Optical light scatter
-Each cell flows in a single line through a
flow cell
-A laser device is focused
-On striking cells, light is scattered in
different directions
-Sensor captures and multiplies scatter
-Forward angle light scatter (FALS) – cell
size
-Side scatter (SS) – granularity
4th aliquot
Delivered to what
Mixed with what
what solution is added
Delivered to a heated dilution chamber
◦ Mixed with new methylene blue reagent
◦ An acidic, hypotonic solution is added
4th aliquot
An acidic, hypotonic solution is added to
◦ Elutes hemoglobin
◦ Preserves precipitated RNA
◦ Spheres RBCs (eliminates interference due to variance
of shape
4th aliquot
Sent to VCS for what
important cell measurement characteristics 1,2,3
◦ Sent to VCS for analysis to classify mature vs.
immature RBCs
1. Cell volume – by impedance
2. Cell conductivity – by electromagnetic probe.
Determines physical and chemical components
3. Cell’s light scatter characteristics - determines internal
contents and cell surface and size
Indicators that may appear
after the data
@
data is outside the linearity limit
*
Data is doubtful
+ or -
Data is outside the reference limits
—– minus minus minus minus
Data doesn’t appear due to analysis error or abnormal sample
++++
Data exceeds display limit
Hematocrit
MCV/ RBC count
MCH
Hb/ RBC count
MCHC
Hb/ Mct
Automated pt count problems
- Plt clumping - recollect in Heparin
- Satellitosis – warm sample
- RBC microcytes
- Giant plts
What to do when controls are out of limit
out of control
stop testing
identify and correct problems
Repeat testing on patient samples and controls
Do not report results until problem is solved and controls indicate proper performance
Coagulation testing
Today, numerous semi-automated & fully automated instruments to perform coagulation testing
Instruments today are likely to have sophisticated front end capabilities
Coagulation testing
Modern congulation analyzers have
reduced what
Random what
Modern coagulation analyzers have
greatly ↑ performance via improved
accuracy & precision
Reduced rgt & sample volumes required
Random, discrete sampling capabilities
Flagging (sample & instrument)
Electromechanical instruments
Instruments
- Fibromter (BBL Microbiology Systems, Becton Dickinson)
Semi-automated: reagents & samples usually added manually by operator
Principle & Operation - Electromechanical clot detection
Stationary & moving probes
Electromagnetic monitoring
- Electromagnetic monitoring of movement
of a steel ball in plasma
Ball moves because of applied magnetic
force
Clot = ↓ movement sensed by electronic
sensor set to predefined limit
Photo optical systems
*Detection of sample optical
density due to formation of fibrin
*Records decreased light at the
forward 180 ̊ angle