Anesthesia Machine 6.1 Flashcards
Identify the components of the low-pressure sytem in the anesthesia machine. (Select 2)
a. flowmeter tubes
b. oxygen flush valve
c. common gas outlet
d. cylinder pressure regulator
anesthesia machine
a. flowmeter tubes
c. common gas outlet
pre-lesson Q1
Match each component to its proper location in the SPDD model.
1. supply
2. disposal
3. delivery
4. processing
a. cylinder
b. vaporizer
c. circle system
d. scavenger
anesthesia machine
1a
2d
3c
4b
pre-lesson Q2
The PISS system is used to:
a. prevent the use of the wrong gas hose
b. prevent the use of the wrong gas cylinder
c. diagram the pathway of the gas through the anesthesia machine
d. filter and exchange the air in the operating room
anesthesia machine
b. prevent the use of the wrong gas cylinder
pre-lesson Q3
The bourdon pressure gauge on an oxygen cylinder reads 500 psi. If the flow rate is 2 L/min, how long will this cylinder provide oxygen to the patient? (minutes, whole number)
anesthesia machine
75, 83, 87 minutes
pre-lesson Q4
Identify the best practices for handling gas cylinders. (Select 2)
a. laying the cylinder on its side when changing the cylinder on the anesthesia machine
b. oiling cylinder valves every 6 months
c. placing two washers between the cylinder and the hanger yoke assembly
d. removing the plastic cover on the port when installing the cylinder
anesthesia machine
a. laying the cylinder on its side when changing the cylinder on the anesthesia machine
d. removing the plastic cover on the port when installing the cylinder
pre-lesson Q5
The oxygen pressure failure device activates when the: (Select 2)
a. oxygen pressure in the supply line is less than 20 psi
b. the oxygen pipeline contains nitrous oxide
c. FiO2 in the inspiratory limb is less than 21%
d. oxygen tank is exhausted
anesthesia machine
a. oxygen tank is exhausted
d. oxygen pressure in the supply line is less than 20 psi
pre-lesson Q6
The hypoxia prevention safety device on the anesthesia machine will:
a. shut off the flow of nitrous oxide if the oxygen supply pressure is less than 30 psi.
b. limit the nitrous oxide flow to three times oxygen flow
c. alarm if the FiO2 is less than 21%
d. will prevent a hypoxic mixture if a third gas is used
anesthesia machine
b. limit the nitrous oxide flow to three times oxygen flow
pre-lesson Q7
Which type of flowmeter float is read at the middle?
a. skirted
b. ball
c. plumb bob
d. nonrotating
anesthesia machine
b. ball
pre-lesson Q8
You are administering air 1 L/min and oxygen 3 L/min. Calculate the fraction of inspired oxygen. (percentage)
anesthesia machine
80
pre-lesson Q9
You are using an anesthesia machine that couples fresh gas flow to tidal volume and fully compensates for circuit compliance. Calculate the total tidal volume delivered to the patient. (mL, whole number)
oxygen: 3 L/min
air: 1 L/min
I:E: 1:2
bellows: 500 mL
RR: 10 bpm
anesthesia machine
632
pre-lesson Q10
A ventilator is programmed to deliver a tidal volume of 600 mL. If the breathing circuit compliance is 5 mL/cm H2O and the peak pressure is 25 cm H2O, what is the total tidal volume delivered to the patient?
a. 425 mL
b. 450 mL
c. 475 mL
d. 500 mL
anesthesia machine
b 475 mL
pre-lesson Q11
The isoflurane dial is set to 2%. What percent of fresh gas exiting the vaporizing chamber is saturated with isoflurane? (percentage)
anesthesia machine
100%
the rest of the fresh gas bypasses the vaporizing chamber
pre-lesson Q12
Which device will be the first to detect an oxygen pipeline crossover?
a. proportioning system
b. oxygen analyzer
c. fail-safe device
d. pulse oximeter
anesthesia machine
b. oxygen analyzer
pre-lesson Q14
What are the most critical actions to perform during an oxygen pipeline crossover? (Select 2)
a. disconnect the pipeline supply
b. replace the oxygen analyzer
c. ventilate with an ambu with the auxiliary oxygen flowmeter
d. turn on the oxygen tank
anesthesia machine
a. disconnect the pipeline supply
d. turn on the oxygen tank
pre-lesson Q15
Select the true statements regarding the oxygen flush valve. (Select 2)
a. the risk of barotrauma is minimzed by ventilators with fresh gas decoupling
b. it will deliver a continous pressure of 35-75 psi
c. excessive use can lead to awareness
d. it will cause the ventilator spill valve to close during inspiration
anesthesia machine
a. the risk of barotrauma is minimized by ventilators with fresh gas decoupling
c. excessive use can lead to awareness
pre-lesson Q16
Select the true statements about the pneumatic ventilator bellows. (Select 2)
a. a descending bellows is made safer by fresh gas coupling
b. a descending bellows cannot rise and fall with a circuit disconnect
c. a leak in the bellows may cause the reading on the oxygen analyzer to increase
d. a hole in the bellows may cause barotrauma
anesthesia machine
c. a leak in the bellows may cause the reading on the oxygen analyzer to increase
d. a hole in the bellows may cause barotrauma
pre-lesson Q17
The piston ventilator: (Select 2)
a. relies on fresh gas flow coupling
b. preserves tank oxygen in the event of oxygen pipeline failure
c. allows for more precise delivery of tidal volumes
d. removes the risk of barotrauma
anesthesia machine
b. preserves tank oxygen in the event of oxygen pipeline failure
c. allows for more precise delivery of tidal volumes
pre-lesson Q18
Which statement regarding pressure controle ventilation is true? (Select 2)
a. the risk of ventilator-associated lung injury is decreased
b. increased lung compliance will decrease tidal volume
c. gas flow decelerates during inspiration
d. the ventilator switches to expiration after a preset pressure is achieved
anesthesia machine
a. the risk of ventilator-associated lung injury is decreased
c. gas flow decelerates during inspiration
pre-lesson Q19
Which modes of mechanical ventilation are best suited for a laryngeal mask airway? (Select 2)
a. synchronized intermittent mandatory ventilation
b. inverse ratio ventilation
c. pressure support ventilation
d. controlled mandatory ventilation
anesthesia machine
a. sychronized intermittent mandatory ventilation
c. pressure support ventilation
pre-lesson Q20
At what pH does the ethyl violet change to purple?
a. 7.5
b. 8.6
c. 10.3
d. 12.1
anesthesia machine
c. 10.3
pre-lesson Q21
When compared to soda lime, what factor is increased with the use of calcium hydroxide lime?
a. carbon monoxide
b. frequency of of replacement
c. CO2 absorption capacity
d. fire risk
anesthesia machine
b. frequency of replacement
pre-lesson Q22
While a patient is ventilating spontaneously with an endotracheal tube, you notice that a fresh gas flow of 10 L/min is required to fill the breathing bag and determine that the scavenger is malfunctioning. Which statement must be true?
a. the negative pressure relief valve has failed
b. the positive pressure relieft valve as failed
c. there is an open scavenging system
d. there is a passive scavenging system
anesthesia machine
a. the negative pressure relief valve has failed
pre-lesson Q23
What is the most common cause of low circuit pressure?
a. leak in corrugated tubing
b. circuit disconnect
c. incompetent ventilator relief valve
d. improper fitting of carbon dioxide absorbent
anesthesia machine
b. circuit disconnect
pre-lesson Q24
During a general anesthetic with an endotracheal tube, the high peak pressure alarm sounds. After changing the bag selector switch from ventilator mode to bag mode, the peak inspiratory pressure returns to baseline. What is the most likely explanation for the rise in peak inspiratory pressure?
a. the positive pressure relief valve on the scavenger failed
b. the endotracheal tube was kinked
c. the patient experienced bronchospasm
d. the ventilator spill valve malfunctioned
anesthesia machine
d. the ventilator spill valve malfunctioned
pre-lesson Q25