7. Radiographic Equipment Flashcards

1
Q

a device that measures the quantity of radiation reaches the image receptor

A

Automatic Exposure Control (AEC)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

it automatically terminates exposure when IR has received that required radiation intensity

A

Automatic Exposure Control (AEC)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

AEC systems also are called

A

automatic exposure devices

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

once a predetermined amount of radiation is transmitted through a patient, the x-ray exposure is terminated

A

Automatic Exposure Control (AEC)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

this determines the exposure time and therefore the total amount of radiation exposure to the image receptor

A

Automatic Exposure Control (AEC)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

2 Types of AEC System

A
  • Phototimers

* Ionization Chamber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

specifically refers to the use of an AEC device that uses photomultiplier tubes or photodiodes

A

Phototiming

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

are the AEC devices that measure the amount of radiation transmitted

A

Detectors

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

used for exposure timer checks

A

Solid-state Radiation Detectors

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

operate with a very accurate internal clock based on a quartz-crystal oscillator

A

Solid-state Radiation Detectors

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

fluorescent (light-producing) screen and a device that converts the light to electricity

A

Phototimers

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

an electronic device that converts visible light energy into electrical energy

A

Photomultiplier Tube (PMT)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

is a solid-state device that performs the same function of a PMT

A

photodiode

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

also called exit-type devices

A

Photomultiplier Tube (PMT)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

flat, parallel plate ______ positioned between the patient and the image receptor

A

Ionization Chamber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

a hollow cell that contains air and is connected to the timer circuit via an electrical wire

A

Ionization Chamber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

considered as entrance-type devices

A

Ionization Chamber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

less sophisticated, less accurate, less prone to failure

A

Ionization Chamber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

interacts with exit radiation before it reaches the image receptor

A

Ionization Chamber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

air in the chamber is ionized and electric charge that is proportional to the amount of radiation is created

A

Ionization Chamber

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

If the radiographic unit has a mAs readout display, the radiographer should observe the reading ____ the exposure is made

A

after

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

The radiographer must be sure to set the ___ value as needed to ensure adequate penetration and to produce the appropriate scale of contrast

A

kVp

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

The ___ selected determines the length of exposure time when using AEC

A

kVp

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

A ___ kVp required more exposure time when using AEC

A

low

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
requires more exposure time to reach the predetermined amount of exposure
low kVp
26
decreases the exposure time to reach the predetermined amount of exposure and reduces the overall radiation exposure to the patient
high kVp
27
If the radiographer can set the __ when using AEC, it affects the time of exposure for a given procedure
mA
28
Increasing the mA _____ the exposure time to reach the predetermined amount of exposure
decreases
29
Decreasing the mA _____ exposure time to reach the predetermined amount of exposure
increases
30
Refers to the shortest exposure time that the system can produce
Minimum Response Time
31
Minimum response time usually is ____ with AEC systems than with other types of radiographic timers
longer
32
If the minimum response time is longer than the amount of time needed to terminate the exposure, an _____ amount of radiation reaches the image receptor
increased
33
Refers to the maximum length of the time the x-ray exposure continues when using an AEC system
Backup Time
34
acts as safety mechanism when AEC System fails or equipment is not used properly
Backup Time
35
Protects the patient from unnecessary exposure and protects the x-ray tube from reaching or exceeding its heat-loading capacity
Backup Time
36
If the back-up time is controlled automatically, it should terminate at a maximum of ___ mAs
600 mAs
37
When in AEC mode, the electronic timer should be set to ___ times the expected exposure time as back-up timer in case the AEC fails
1.5 times
38
Backup time should be at _____ of the expected exposure time. This allows the properly used AEC system to appropriately terminate the exposure, but protects the patient and tube from excessive exposure if a problem occurs
150% to 200%
39
To _____ patient exposure, the backup time should be neither too long or too short
minimize
40
The combination of _____ affect the amount of exposure reaching the image receptor
detectors
41
If the area of radiographic interest is not directly over the selected detectors, that area probably will be
over-or underexposed
42
When performing any radiographic study in which the image receptor is located outside of the Bucky, the ____ should be deactivated and a manual technique used
AEC system
43
Accurate centering of the area of interest over the detectors is ____ to ensure proper exposure to the image receptor
critical
44
If the area of interest is not properly centered to the image receptor, ______ may occur
over-or underexposure
45
The ___ of the detectors used within an AEC system is fixed and cannot be adjusted
size
46
If the detector combination is larger than the area of interest, a ____ exposure technique should be used
manual
47
____ factors affect the time the exposure takes to reach the image receptor and ultimately affect image quality
Patient
48
Variations in patient _____ result in changes in the time of exposure accordingly if AEC system is functioning properly
thickness
49
Pathologic conditions, contrast media, foreign objects, or pockets of gas are ______ that may affect the proper exposure to the IR and ultimately image quality
patient variations
50
Excessive or insufficient _____ may affect the amount of exposure reaching the image receptor
collimation
51
Insufficient collimation may result in excessive ____ reaching the detectors, resulting in the exposure time terminating too quickly
scatter
52
Excessive collimation may result in an exposure time that is too ___
long
53
is calibrated to the type and speed of the IR used
AEC System
54
If an image receptor of a different type or speed is used, the detectors will not sense the difference, the exposure time will terminate at the preset value, and image quality may be _____
jeopardized
55
Refers to a radiographic system that allows the radiographer to select a particular button on the control panel that represents an anatomic area; a pre-programmed set of exposure factors can be selected
Anatomically Programmed Radiography (APR)
56
are not related in their function
APR and AEC
57
The purpose of ______ is to ensure that consistent and appropriate exposures to the image receptor are produced
calibration
58
Failure to maintain regular _____ of the unit results in the lack of consistent and reproducible exposures to the detectors and could affect image quality
calibration
59
is important in monitoring the performance of the AEC system
Quality Control
60
Optical densities should be within ____ and reproducibility of the exposures for a given set of exposure factors and selected detector should result in mR readings within __ and optical densities within ____
+/- 0.2, 5%, +/- 0.10