UNIT 3: Equipment Design Flashcards
Primary/ useful beam
those x-rays emitted through the x-ray port, or tube window
Scattered radiation
all the non-useful-image-forming radiation that arises from the interaction of an x-ray beam with the atoms of a patient or any other object in the path of the beam
Off focus radiation
X-rays emitted from parts of the x-ray tube other than the focal spot. Also called stem radiation
PBL
(Positive beam limitation) A feature of radiographic collimators that automatically adjusts them so that the radiation field size matches the size of the image receptor. Also known as automatic collimation
Filtration
Elements that are part of or added to the x-ray tube to reduce exposure to the patient’s skin and superficial tissue by absorbing most of the lower energy photons from the produced heterogeneous beam and thereby increasing the mean energy, or quality, of the x-ray beam
Total filtration
inherent filtration and added filtration jointly combine to equal the required amount necessary to filter the useful beam adequately (2.5 mm AI/Eq)
HVL
(Half-value layer) The thickness of a designated absorber (customarily a metal such as aluminum) required to decrease the intensity of the primary x-ray beam by 50% of its initial value
Exposure Reproducibility
Consistency in output in radiation intensity for identical generator settings from one individual exposure to subsequent exposures. This means that the x-ray unit must have the ability to duplicate certain radiographic exposures for any given combination of peak kilovolts (kVp), milliamperes (mA), and time. A variance of 5% or less is acceptable.
Exposure Linearity
Consistency in output radiation intensity at any selected kVp when x-ray generator settings are changed from one milliamperage and time combination to another. Mathematically, it is the ratio of the difference in mR/mAs values between two successive x-ray unit generator stations to the sum of those mR/mAs values. It must be less than 0.1. When changing from one mA station to a neighboring mA station, the most that linearity can vary is 10%
HLC
(High Level Control) An operating mode for state-of-the-art fluoroscopic equipment in which patient entrance exposure rates are substantially higher than normally allowed for routine procedures. The higher exposure rate allows visualization of smaller and lower contrast objects that do not usually appear during standard fluoroscopy. HLCF is also known as “boost” mode
How is the protective tube housing constructed to protect the patient and radiographer?
The housing enclosing the x-ray tube must be constructed with lead lining to reduce radiation leakage through any portion of the housing away from the useful beam
What is the maximum allowable leakage from the x-ray tube? (inc. distance)
.88 mGy/hour at 1 meter from the source (when operated at highest tube voltage and current)
What radiation protection requirements are related to the console/ control panel?
For the radiation safety of the radiographer, it must be located behind a properly shielded barrier that also has a radiation-absorbent window, Fully display the conditions of exposure and provide a positive indication when the x-ray tube is energized, exposure hand control be mechanically affixed to the console such that it cannot be activated while the operator is in an unshielded location.
Identify the optimal characteristics of the radiographic exam table
support patients whose weight can be up to 400 pounds, a free-moving tabletop that allows easy movement of the patient during an imaging procedure, thickness of the tabletop must be uniform, for under-table x-ray tubes as used in fluoroscopy, the patient support surface should be as radiolucent as possible, Carbon fiber commonly used.
What is the recommended relationship between collimation and the IR?
Tape measure or laser Distance must be accurate within 2% of SID, Light localizer to be within 2% of SID, Centering indicator must be accurate to within 1% of SID. The primary x-ray beam shall be adequately collimated so that it is no larger than the size of the image receptor being used for the examination
Describe the optimal construction characteristics of the collimation system
*Two sets of adjustable lead shutters mounted within the device at different levels
*A light source to illuminate the x-ray field and permit it to be centered over the area of clinical interest
*A mirror to deflect the light beam toward the patient to be imaged
How far should the patient’s skin be from the collimator for fixed radiographic equipment?
At least 15 cm(6 inches) below the collimator
How far should the patient’s skin be from the collimator for portable or mobile radiographic equipment?
at least 30 cm (12 inches)
What is the purpose of the light in the x-ray tube housing? What is the relationship between the light and the primary beam?
-to outline the desired margins of the radiographic beam adequately on the patient’s anatomy
-Good coincidence (i.e. very similar physical size and overlapping alignment) between the x-ray beam and the localizing light beam is essential to eliminate collimator cutoff of the body structures that need to be imaged
How would a variance greater than 2% between the beam/ light source affect radiation protection?
The beam would be out of Alignment
What would happen if the PBL and IR agreements were not in sync?
The positive beam limitation (PBL) is to restrict the size and shape of the x-ray beam so it does not exceed the size of the selected image receptor, if they’re not in sync areas of the patient outside the image receptor would be irritated
What is the relationship between cones, flared metal tubes and straight cylinder?
Cones are circular metal tubes that attach to the x-ray tube housing or variable rectangular collimator to limit the x-ray beam to a predetermined size and shape. Cones can be Flared metal tubes or Straight cylinders
What is the purpose of filtration?
Reduces exposure to patient’s skin and superficial tissues by absorbing low energy photons (long-wavelength or soft x-rays)
How does filtration impact the characteristics of the primary x-ray beam?
Increases quality or mean energy of beam
Describe how filtration reduces absorbed dose (D) to the patient.
the absorbed dose to the patient decreases because very low-energy photons would increase the patient’s radiation dose if it weren’t for filtration
Inherent filtration includes:
Glass envelope encasing x-ray tube
Insulating oil surrounding x-ray tube
Glass window in tube housing