Tono 3 Flashcards

1
Q

Applanation instruments

A

Goldmann
Perkins
Tonopen (microstrain gauge)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Dynamic contour instrument

A

Pascal

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Rebound instrument

A

iCare

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Indentation instrument

A

Non contact (puff) , Contact (Shiotz)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Pressure phosphene instrument

A

Proview

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Transpalpebral instrument

A

Diaton

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Tono methods (6)

A
Applanation (includes microstrain gauge) 
Dynamic Contour 
Rebound
Indentation
Pressure phosphene 
Transpalpebral
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Indentation method

  • Measures what?
  • Two types
A
  • Measures amount of force required to indent entire cornea

- Non contact (air puff) or contact (shiotz)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Applanation technique

  • Measures what?
  • Examples
A

Measures amount of force required to flatten the corneal surface only.
Goldmann, Perkins, tono

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Tonopen

  • method
  • useful when
  • Measurement affected by what?
  • Tends to measure low or high?
  • Statistical coefficients
  • Measurement range
A
  • Applanation technique with microstrain gauge
  • Useful in children, supine patients, gross corneal path
  • Affected by central corneal thickness
  • Tends to measure high
  • 5% is the best, 10%, 20%
  • range is 5-80 mmHg (less accurate at higher IOPs)

Can use on peripheral cornea
Consider latex allergies

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Main differences between ton open and ton open avia

A

Avia is smaller and averages 10 readings instead of only 3-4.
Larger LCD display
Measurement range is lower: 5-55 mmHg. Pen is 5-80 mmHg.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Perkins applanation tono

  • Advantages
  • Disadvantages
  • Pressure range
A

Advantage: Portable, good for supine patients
Disadvantage: Fixed mag, cannot examine cornea before and after, use anesthetic.
Range: 0-80

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Dynamic contour tono (pascal)

A

Slit lamp mounted instrument
Very expensive
Independent of central corneal thickness !!!!!
Measures cardiac cycle
Average reading given with a Q value for quality
Requires several seconds of steady fixation
Range 5-80
calculated and displayed on screen

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

i care rebound tono range and pros/cons

A

range is 7-50
pros: no anesthesia required
Cons: Can be influenced by CCT, over estimate IOP by 0.6-1.8

  • can get i cares at home to measure by self.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

NCT range

Pros and cons

A

Range is 0-60
Pros: easy to use, no anesthetic
cons: Less reliable than goldmann. Overestimates. Can be affected by apprehension, keratometry and cardiac cycles.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Pressure phosphene tono (PPT)

How do you admin test

A

Pressure is applied to the patients globe on eyelid by pressing the plunger. Have patient look down and out. Once the pressure in the globe equals the pressure applied by probe, the patient will see a phosphene- circular or arc shaped light.

17
Q

pressure phosphene tono (PPT)
pros/cons
range

A

pros: good for anyone who can report seeing the phosphene
Cons: Not relibable, must have compliant patient, limited scale
8-40

18
Q

Transpalpebral tono how to admin

A

Set patient fixation at a 45 degree angle
Position rod on lid and perform 5-6 x then average.
It calculates the eyeballs resistance at cornea and converts to IOP.

Ranges 5-60

19
Q

Corneal hysteresis

What is it and what are the ranges

A

How well the cornea can absorb energy and dissipate it. Also depends on IOP.

Less than 8mmHg= Risk for glaucoma. K can’t take impact well and ON will likely have same properties.

Medium is 8-12 mmHg

High is 12+ mmHg = good. Less glaucoma risk.