Test #1 Terms, Machine, Form Flashcards
The maintenance period is achieved by use of 1)___________ or 2)_____________ anesthetics. We want to see a 3)__________ increase of 4)_________ rate and 5)__________ rate. We must guard against 6)____________ anesthesia.
1.injectable 2.inhalant 3.slight 4.heart 5.respiration 6.chasing
T/F Epinephrine maybe added to a local anesthetic agent to prolong the effects of the drug
True
If a precision vaporizer is VOC, then a non-precision vaporizer is ______
VIC (Vaporizer-in-circuit)
During the anesthetic procedure, the reservoir bag should always be about _____ full of air
3/4
What is the negative pressure relief valve for?
It is a safety device only. It allows room air to enter the circuit.
Steps to pressure check
- Close pop-off 2. Plug end of Y tubing with thumb 3. Flush valve/ Flow meter up to 20cm H2O 4. Watch pressure manometer (must not drop >5cm H2O in 30 seconds) 5. DO NOT release thumb-> open pop-off to release the pressure in the circuit 6. Turn OFF flow meter
T/F An epidural is a common example of regional anesthesia
True
T/F Intravascular routes of anesthesia have a longer duration than intramuscular
False Shorter
T/F The pressure regulator will reduce the pressure of oxygen entering the anesthetic machine to 50 psi
True
T/F The flowmeter should never be set at less than 2L/minute when running anesthesia on a patient that weighs less than 7 kg
False
T/F It is not important to turn the oxygen cylinder off when the anesthesia machine is not in use
False It is important
T/F 600 minutes will be left on E tank of oxygen containing 300 L at a flow rate of 500 ML/minute
True
T/F Neuroleptanalgesia is produced by a combination of a narcotic opioid and a tranquilizer or sedative
True
T/F Rebreathing systems are used on patients that way less than 7 kg
False Patients greater than 7 kg
T/F Heart rate, respiratory rate, and blood pressure are recorded every 15 minutes on the bel-rea anesthesia form
False Every 5 min