Shielding Flashcards

1
Q

Large rooms in NM serve to

A

increase distance and reduce exposure to other staff and other equipment

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

The NM facility design should be or contain

A

remote from the main hospital busy areas

connected with imaging areas

have good access for deliveries and patients leaving

separate patients and staff

imaging rooms

patient reception and waiting

hot lab

radionuclide storage

patient injection area

waste storage

cardiac stressing area

air extraction for ventilation studies

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Typical lead shielding for CT

A

1/16”

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Half-value layer for lead for Tc-99m

A

.3 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Half-value layer for lead for F-18

A

5.5 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

1/16” is approximately

A

1.6 mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly