Gamma Cameras and SPECT Flashcards
The resolution of a collimator is expressed as the
FWHM of a line source
Collimators generally involve a tradeoff between
spatial resolution
sensitivity
parallel-hole collimators may be classified as
high sensitivity (low resolution)
general-purpose
high resolution (low-sensitivity)
Typical collimator for Tc-99m and Tl-201
low energy
Typical collimator for Ga-67, In-111
Medium energy
Typical collimator for I-131 and F-18 (non-PET)
High-energy
Typical count range for planar image
500,000 to 1 M
SPECT images are typically reconstructed using this matrix
128 x 128
For brain SPECT, typical collimators may include
hybrid convergin and parallel hole
Typical bit depth in terms of gray levels
8 bit, 256 gray levels
Gamma cameras typically have this number of PMTs
37, 61, 91 PMTs
A typical window for Tc-99m imaging
20%
Efficiency of the detector crystal increases with
thickness
Inclusion of scattered photons in a gamma camera acquisition causes
decrease in spatial resolution
Primary cause of limited intrinsic resolution in a gamma camera
statistical fluctuation in the distribution of light photons among photmultiplier tubes from one scintillation to the next
As gamma or x-ray energy decreases, intrinsic resolution
decreases do to larger relative statistical fluctuations in the light photons per scintillation event.
As a rule of thumb, intrinsic resolution in a gamma camera and energy are related as
1 / √E
(similar to relative uncertainty in count statistics)
Thicker detectors result in
increased sensitivity
decreased intrinsic resolution
The system resolution is largely determined by the
collimator resolution
Efficiency for 140 keV with NaI(Tl)
70% to 90% for 6.4mm and 12.7 mm thick crystals
Efficiency of NaI(Tl) for 500 keV photons
less than 20%
The energy resolution improves according to photon energy as
1 / √E
due to relatively decreased statistical fluctuations at higher energy
For Tc-99m, a lower energy threshold of 130 keV should remove photons that have been scattered through angles greater than
45 degrees, but only at about 50% due to the blurring of the spectrum (for scattered and unscattered photons)
As count rates increase, the accuracy of dead time corrections
decreases
True or False
In NM imaging studies,most are not impaced by dead time losses
True, with the exception of high-counting rate applications such as first-pass cardiac studies
In a NM image with high count rates from two sources, the resulting image may look like
An undesirable effect of pulse pile-up in terms of the image
image distortion
Modern gamma cameras employ this approach for pile-up correction
pulse-tail extrapolation
Which happens between the PMTs and over the PMTs?
Pincushion occurs over the PMT due to nonlinearities in sensitivity as you approach the center. Barrel distortion occurs between PMTs
Cause of artifacts
nonlinearities: pincushion and barrel effects seen with line pattern and homogeneity test
Two causes of gamma camera nonuniformities
- Differences in pulse-height spectrum for different PM tubes (tube response can be adjusted with careful tuning)
- position-dependent collection efficiency of scintillator light
Edge packing, due to reflection of scintillator light from the sides of the detector crystal back towards the PM tubes is mainly masked by
Useful Field of View
UFOV
For higher energy photons, interactions occur closer to the PMTs, resulting in a more narrow light spread and a __________ of detector nonuniformity
worsening
An intrinsic flood image is used to apply
spatially varying energy corrections
spatially varying energy corrections include
- spatially varying PHA settings
- normalized intensity correction (count skimming) : certain percentage of counts in hot spots are thrown out {older approach}
- use of lead sheet with uniformly spaced array of holes (~1mm diam, ~4 mm separation) at known locations, placed without collimator - lookup table corrects for offsets in x and y {mfg}