A dynamic series of analogue consecutive intrinsic spatial resolution images using 99mTc were made to check the consistency of the image formatter. The energy window had been centred over the photopeak of 99mTc prior to the check.
energy window over the photopeak has shifted, resulting in patchiness
Tc DTPA of kidney
Poor energy window position, note appearance of PM tubes
good uniformity image, intrinsic, 3M counts, Tc - 99m
septal penetration
streak artifacts appear along directions where septa is thinnest and can extend over many cm indicating penetration of many holes
In this image a high energy collimator with hexagonal holes was used
PMT voltage drift causing peak shift and difference in sensitivity
High count rates can lead to a loss of
Intrinsic flood image
crushed lead septa; lead foil separation
cracked crystal (mechanical shock or thermal shock - when changing temp by more than 10 deg / hr)
truncation artifact
Which set of images was performed with a 360 vs a 180 degree acquisition arc?
bottom
Which is a low resolution SPECT image, 60 images at 6 degree steps or 120 images at 3 degree steps?
60 images at 6 degree steps
The typical number of counts in a spect study range from
2 - 8 M
To reduce high frequency noise in a FBP image, use a
low pass filter such as hannning or butterworth