Artifacts and Images Flashcards

1
Q
A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

A dynamic series of analogue consecutive intrinsic spatial resolution images using 99mTc were made to check the consistency of the image formatter. The energy window had been centred over the photopeak of 99mTc prior to the check.

A

energy window over the photopeak has shifted, resulting in patchiness

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Tc DTPA of kidney

A

Poor energy window position, note appearance of PM tubes

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q
A

good uniformity image, intrinsic, 3M counts, Tc - 99m

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q
A

septal penetration

streak artifacts appear along directions where septa is thinnest and can extend over many cm indicating penetration of many holes

In this image a high energy collimator with hexagonal holes was used

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q
A

PMT voltage drift causing peak shift and difference in sensitivity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

High count rates can lead to a loss of

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Intrinsic flood image

A
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q
A

crushed lead septa; lead foil separation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q
A

cracked crystal (mechanical shock or thermal shock - when changing temp by more than 10 deg / hr)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q
A

truncation artifact

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Which set of images was performed with a 360 vs a 180 degree acquisition arc?

A

bottom

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Which is a low resolution SPECT image, 60 images at 6 degree steps or 120 images at 3 degree steps?

A

60 images at 6 degree steps

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

The typical number of counts in a spect study range from

A

2 - 8 M

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

To reduce high frequency noise in a FBP image, use a

A

low pass filter such as hannning or butterworth

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q
A

bulls eye artifact due to poor uniformity

17
Q
A

serial ring artifact

18
Q
A

COR artifact