scavenging systems Flashcards

1
Q

NIOSH volatile anesthetic alone

A

2ppm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

NIOSH nitrous oxide

A

25ppm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

NIOSH volatile agent w NP

A

0.5ppm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

5 components of scavenging system

A
1 - gas collecting assembly
2 - transfer means
3 - scavenging interface
4 - gas disposal tubing 
5 - gas disposal assembly
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

gas collecting assembly

A
  • captures gases at site of emission
  • delivers to transfer means tubing
  • outlet connection usually 30mm (19 older) male
  • size of connection important so that it doesn’t connect to other components of breathing system
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

transfer means

A
  • “exhaust tubing” … “hose” … “transfer system”
  • convery cad from collecting assembly to interface
  • usually tube w female on both ends
  • tubing short
How well did you know this?
1
Not at all
2
3
4
5
Perfectly