we need to top the reservoir up, it can obly be filled with purified water (AWD)
T or F
ture
If it’s still wet after sterilisation we can remove the moisture with a sterilised cloth.
T or F
false
Contamination was identified at inspection but it is alright to wipe it off as its getting sterilised.
T or F
false
As long as we clean them between times, we can transport our instrumentation in any available container.
T or F
false
Operator doesn’t need PPE if they are only using the Ultrasonic
T or F
false
As the Ultrasonic is a powerful process there should be no need to further process the instrumentation, processed through this equipment.
Tor F
false
The standard disinfection temperature banding within our washer-disinfector is
Select one:
a. 80 - 85
b. 85 - 90
c. 75 - 85
d. 65 - 75
e. 90- 95
e
90-95
identify stage A in AWD
flush
identify stage B in AWD
wash
identify stage C in AWD
rinse
identify stage D in AWD
disinfection
identify stage E in AWD
drying
Guidance documents are just there for reference, I don’t need to follow the directions within them
T or F
false
The first stage of a vacuum capable or Type - B steriliser is
Select one:
a. Equilibration
b. Steam penetration
c. Air removal
d. Negative pulsing
e. Heat up
f. Positive pulsing
c
air removal
A Bowie-Dick is carried out a daily basis, What is it testing for
Select one:
a. Temperature
b. Steam penetration
c. Air removal
d. Presence of air
e. Steam dryness
b.
steam penetration