quality control/calibrations Flashcards
helium dilution calibration
2 or 3-point done before every test
point1= 0% Helium
point2/3= known concentrations of Helium
nitrogen washout calibration
2-point once every 6 months
point1= 78% N2 (room air)
point2= 0% nitrogen
sanz electrode
pH
severinghaus electrode
PCO2
clark electrode
PO2
partial pressure of end tidal CO2
PetCO2 significantly lower than PCO2 indicates
- embolus
- decreased cardiac output
- cardiac arrest
PetCO2 calibration
capnography
requires two concentrations
- zero calibration on room air
- 5% CO2
blood gas 1-point calibration
- adjust analyzer to single standard
2. before sample analysis or every 30 minstrels
blood gas 2-point
- adjust analyzer to two standards
2. every 2-8 hrs
pH calibration
one-point: 7.38
two-point: 7.38 & 6.84
CO2 calibration
one-point: 5%
two-point: 5% and 10%
02 calibration
one-point: 0%
two-point: 0% & 10% 12%
clark electrode O2 analyzer
- requires batteries
2. fast response
galvanic O2 analyzer
- uses a fuel cell
2. slow response
O2 analyzers calibration
2-point on 100% O2 and room air(21%)