QUALITY CONTROL AND ASSURANCE Flashcards
What are the two main areas of activity for quality control in radiology?
Quality Assurance (QA) and Quality Control (QC)
What organization promotes the “Ten-Step Monitoring and Evaluation Process”?
The Joint Commission (TJC)
What is the main focus of Quality Assurance (QA)?
People, including patient scheduling, reception, and preparation
Who is responsible for a QA program?
The radiologist
What does Quality Control (QC) primarily focus on?
Instrumentation and equipment
Who is responsible for a QC program?
Medical physicist
What is the first step in the TJC’s Quality Assurance Program?
Assign responsibility
What is the purpose of routine performance monitoring in QC?
To ensure optimal image quality with minimal patient radiation exposure
What is Acceptance Testing in QC?
Testing new radiologic equipment before clinical use
What does preventive maintenance in QC aim to do?
Make repairs unnecessary by regular maintenance
What is the minimum total filtration required for x-ray beams?
2.5 mm Al
How often should collimation be checked?
Semiannually
What should the focal-spot size accuracy be within?
± 50%
What is the tolerance for kVp calibration accuracy?
± 10%
What devices are used to measure kVp calibration?
Filtered ion chamber and photodiodes
What is the frequency of exposure timer accuracy testing?
Annually
What is the acceptable accuracy for exposure times greater than 10 ms?
± 5%
How often should exposure linearity be tested?
Annually
What should exposure reproducibility be within?
± 5%
What is the function of Positive Beam-Limiting (PBL) collimators?
Automatically adjust the x-ray field size to the image receptor size
What is used to measure focal-spot size?
Pinhole camera, star pattern, and slit camera
What is a limitation of the star pattern method for focal-spot size measurement?
Ineffective for sizes less than 3 mm
What does kVp calibration ensure?
The x-ray generator produces the correct energy levels
What happens if the kVp varies by 4%?
It affects image optical density and contrast
What is the purpose of Automatic Exposure Control (AEC)?
To provide a constant image density regardless of tissue differences
How is exposure linearity determined?
Using a precision radiation dosimeter
What should the accuracy of section level in tomographic QC be within?
± 2 mm
What is the required minimum half-value layer (HVL) for a single-phase x-ray unit at 70 kVp?
1.6 mm Al
What should processor cleaning frequency be?
Weekly