QBC Star Instrument Flashcards
What are the two routine user controls on a QBC Star?
- “Star” button
- Door release latch
What are the functions of the “Star” button?
- Teal colored
- Start testing
- Abort testing
- Reprint test results
How often is it recommended to turn off QBC Star?
At the end of the day
What is a principle of testing of the QBC Star?
Makes it possible to quantify buggy coat causing separation of cells by spinning blood at a high rate of speed
What are the materials and equipment required for testing?
- Whole blood (EDTA)
- QBC Star collection tubes
- QBC Star Instrument
- Printer paper
How many times do you mix the blood sample immediately before filling the QBC Star tube?
12-15 times
Within what timeframe do you want to place the QBC Star tube into the instrument after filling?
within 15 minutes
How many times do you rock the QBC Star tube to mix blood with orange fluid?
4 times