QA/QC Flashcards
1
Q
- QA Deals with:
- QC Deals with:
- QA stands for:
- QC Stands for:
A
- people
- equipment
- Quality Assurance
- Quality Control
2
Q
- QA involves _______________________
- QA monitors ____________________
- QA responsibility belongs to _____________
- What is the Joint Commission’s 10-step quality assurance program?
A
- Assessment and evaluation of patient care
- patient scheduling, reception & preparation
- attention from all imaging staff but belongs to radiologist & imaging service management
- ensures health care organization will provide high-quality service & patient care
3
Q
- QC involves _____________________
- What is essential in QC?
- New equipment should be ____________
performed by _________ - Who is responsible for QC program?
A
- Set of activities designed to monitor & maintain systems that produce a product
- Documentation
- acceptance tested
- someone other than manufacturer rep - Medical Physicist with help from QC techs & radiologic engineers
4
Q
- 3 steps to QC program are:
- Describe each step:
A.
B.
C.
- Describe each step:
A
- A. Acceptance Tested
- preformed by someone other than manufact. rep. before clinically applied
B. Periodic Monitoring
- with use performance usually decreases, this is done to equipment periodically
C. Preventative Maintenance
- When periodic monitoring shows error, maintenance or repair is needed ( this step prevents need for repairs)
5
Q
- Filtration measured in ______
- Is it measured directly? - Amount of filtration in general x-ray:
- What is filtration?
- What is X-Rays HVL?
- Diagnostic Range?
A
- Half-Value Layer (HVL)
- not possible to be measured directly
- 2.5 mm Al
- Reduction of number of low energy x-rays that dont contribute to image, (they only increase pt dose unnecessarily)
- Measurement of quality of xrays *
*it is thickness of absorbing material (amount of filtration) needed to reduce intensity of xray to half original value- 3-5 mm Al in xray
6
Q
- What is the minimum total filtration required for stationary (fixed) x-ray equipment?
- What is it for fluro?
- If the x-ray field & light field of collimator are misaligned, what can occur?
- What is PBL?
A
- Above 70 = 2.5 mm Al
50-70 = 1.5 mm Al
Below 50 = 0.5 - 2.5 mm Al
- Missed anatomy or unintended anatomy irradiated
- Positive Beam Limitation (automatic collimation)
- doesnt allow field size exceed IR size
7
Q
- Misalignment must not exceed ______
- What is the Penny Test?
- What tools measure the focal spot size?
- How often is it evaluated?
A
- 2% of SIDE
- Light field / xray congruency test
- Slit camera, pinhole camera & star pattern
- Annually or when equipment is replaced
8
Q
- Centering must be accurate to ______
- _______ - standard measurement tool of effective focal spot
- Other focal spot size tool:
- Which is limited? by what?
- What is effective focal spot?
A
- 1% of SIDE
- Slit Camera
- Pinhole Camera & Star Pattern
- Star Pattern limited for focal-spot sizes less than .3mm - Area projected onto patient and IR
9
Q
- 40 SID misalignment must not exceed ____in
- 72: ____in
- Small focal spot sizes used for ____ kvp
- large focal spots used for ____ kvp because:
- Measured kVp should be within ______ of indicated
- or _____
- kVP calibration tool:
- How often checked
A
.8in
- 1.4in
2. smaller kvp / small extremities
- larger kvp/extremities because more heat generated at anode
- 10% indicated
-OR +/- 4 - Digital kVp meter
- annually or when replaced/fixed
10
Q
- Exposure time responsible for ___ & _____
- Accuracy must be within _______ of ___________
- For AEC, if exposure time fails ________ should terminate exposure
- Above terminates at _____ or ____, whichever is first
A
- patient dose and image density
- within 5% of indicated time for exposures greater than 10ms
- Back-Up Timer
- 6s or 600 mAs
11
Q
- ______ method used to check x-ray timer accuracy
- operating on ______ or _______
- With above method, radiograph shows ____ for _________ during exposure time
- _________ used for three-phase equipment
- displays as:
- Radiation intensity should not vary by more than ___
A
- spinning top
- single phase half or full wave rectification - DOT for each pulsation
- Sychronous spinning Top
- Solid Arc - +/- 5%
12
Q
- Exposure Reproducibility Measures:
- Done By Doing:
- Variation Within: - Exposure Linearity Measures:
- Done By Doing:
- Variation within: - How often is protective apparel checked?
- How? - X-Ray Tube Rating Charts are used for:
- 3 Main Charts:
A
- Consistency in output radiation intensity
- same technical exposure factors repeatedly and seeing how it varies
- 5% (use same technique over and over and it must not vary more than 5%) - Ability of x-ray unit to produce constant, steady radiation output at a given mAs with varying mA x Time
- Setting changed rom one mA to neighboring mA
- Within 10% between adjacent mA stations ( mA & s is changed, mAs stays same / must not vary more than 10%) - Annually
- fluro’d or radiographed - Guide Tech. in use of x-ray tubes (every tube has own)
- Radiographic rating chart, anode cooling chart & housing cooling chart
13
Q
- What do radiographic rating charts convey?
- Are they the same for every x-ray tube?
- How to tell on chart which are safe & unsafe?
- Unsafe exposures result in:
A
- which techniques are safe or unsafe for tube operation
- Vary for everyone based off tube characteristics
- Below line is safe, above curve is unsafe
- Tube may fail abruptly
14
Q
- Anode Cooling Chart Determine: (2 things)
- Thermal Energy Measured in:
- Formula:
- Housing Cooling Chart Determines: (2 things)
- Standard image test patterns in Digital Display QC are:
A
- Max heat capacity of anode / heat dissipation
& length of time required for complete cooling after level of heat input - Heat Units (HU) or Joules (J)
- HU = kVp x mA x s - SAME as anode cooling but for housing
- SMPTE Test & AAPM TG 18 Test
15
Q
- What is maximum anode heat capacity?
- How long until anode cools completely?
- If an exam results in delivery of 50,000 HU to anode in matter of seconds, how long until anode cools completely?
A
- 350,000 (peak/highest point on chart)
- 15 min (lowest point on chart / where it goes to 0)
- 9 min
(50,000 HU intersects anode cooling curve at 6min.
- Min Time - Intersection = Cooling Complete Time
- 15 - 6 = 9 )