Which imaging plane can produce a longitudinal image of the IVC and the aorta on the same image
a) mid sagittal only
b) transverse only
c) coronal and transverse
d) coronal only
d) coronal only
You are scanning a gallbladder and there appears to be sludge within the lumen. Which of the following aid in differentiating sludge from artifact
a) increase the transducer frequency
b) ask the patient to fast for 3-4 more hours and rescan the GB for changes
c) ask the patient to perform the valsalva maneuver
d) apply color Doppler
a) increase the transducer frequency
Harmonic imaging will
a) improve temporal resolution
b) reduce the appearance of microbubble contrast on a contrast echo
c) degrade lateral resolution and improve axial resolution
d) allow for deeper penetration with improved spatial resolution
d) allow for deeper penetration with improved spatial resolution
What system control should be turned off when trying to differentiate a small stone in the gallbladder from a sludge ball
a) harmonic imaging
b) focal zone
c) time gain compensation
d) compound imaging
d) compound imaging
What determines the transducer frequency used for the evaluation of a foreign body
a) depth of the foreign body
b) age of the foreign body
c) composition of the foreign body
d) the radiologist reading the exam
a) depth of the foreign body
When evaluating a liver transplant, you must use angle correction during Doppler evaluation of
a) hepatic arteries and portal veins
b) portal veins only
c) hepatic arteries and veins
d) hepatic arteries only
a) hepatic arteries and portal veins
No flow is detected while performing a Color Doppler evaluation of a normal splenic vein. What will improve the color display
a) increase the color velocity scale
b) increase the color packet size
c) increase the sample size
d) increasing the angle of insonation
b) increase the color packet size
A patient presents for a liver ultrasound with contrast to evaluate a tumor. Which of the following lists the correct machine settings needed for the exam
a) mechanical index over 1.5, harmonics off, and focal zone at the abdominal wall
b) highest output power setting available, harmonics on, and at least three focal zones surrounding the area of interest
d) low mechanical index, harmonics on, and focal zone place at the area of interest
d) low mechanical index, harmonics on, and focal zone place at the area of interest
What change in system controls can reduce the appearance of clutter on a color Doppler image
a) increasing persistence
b) increasing color threshold
c) decreasing color priority
d) decreasing color PRF
c) decreasing color priority
In order to assess low velocity flow, which of the following operator adjustments should be made
a) decrease wall filter setting and compression
b) increase scale setting
c) decrease compression
d) decrease the rejection and wall filter setting
d) decrease the rejection and wall filter setting
Which of the following techniques will lead to degraded lateral resolution
a) increased probe frequency
b) increase in number of focal zones
c) none of the above
d) trapezoidal field of view on linear array
d) trapezoidal field of view on linear array
Which of the following correctly describes why the arcuate arteries are not easily or accurately evaluated by PW Doppler
a) because the blood flows parallel to the Doppler sound beam
b) because the flows perpendicular to the Doppler sound beam
c) because the flow is very low resistance
d) the renal capsule attenuates the Doppler signal so only limited information is reflected to the probe
b) because the flows perpendicular to the Doppler sound beam
______ occurs posterior to highly attenuating structures
a) refraction
b) enhancement
c) speckle
d) shadowing
d) shadowing
The use of multiple focal zones will improve_______ but degrade_____
a) lateral resolution, temporal resolution
b) temporal resolution, lateral resolution
c) axial resolution, lateral resolution
d) lateral resolution, axial resolution
a) lateral resolution, temporal resolution
Which of the following will increase the amount of color displayed in the abdominal aorta
a) decrease color threshold
b) decrease color gain
c) increase sample size
d) increase frame rate
d) increase frame rate