PHEP_26B Flashcards
Thermal instruments are used to measure: a) DC b) AC c) heat d) none of the above
c) heat ref: p. 10
Most meter mechanisms are designed to take advantage of two of the effects of current in an electric circuit: a) magnetism-voltage b) heat-magnetism c) heat-voltage d) resistance-heat
b) heat-magnetism ref: p. 45
The type of meter or instrument using the heat-producing effect of current is called a: a) watt instrument b) thermal instrument c) current potential d) voltage instrument
b) thermal instrument ref: p. 45
A wattmeter measures: a) power factor b) current potential c) current-resistance d) current-voltage
d) current-voltage ref: p. 20
Most watt meters are a/an: a) D’Arsonval meter b) electrodynamometer c) potential-coil meter d) iron-vane meter
b) Electro-dynamometer ref: p. 20
The D’Arsonal electromagnetic movement: a) is suitable for either DC or AC measurements b) is used only for AC measurements c) can be converted to read AC by using a bridge rectifier unit in the coil circuit d) is not affected by the magnetic effects of AC
c) can be converted to read AC by using a bridge rectifier unit in the coil circuit ref: p. 5
A copper-oxide rectifier is a device that: a) converts DC to AC b) offers high opposition to current in two directions c) is classed as a uni-directional conductor d) has two good conductors of electricity
c) is classified as a uni-directional conductor ref: p. 5
AC voltages in the higher ranges are usually measured with a: a) voltmeter and potential transformer b) voltmeter and current transformer c) multiplier and voltmeter d) millivolt meter and a shunt in parallel
a) voltmeter and potential transformer ref: p. 16
The range of an AC ammeter may be extended by using a: a) condenser b) straight shunt c) current transformer d) all three of the above
c) current transformer ref: p. 14
AC ammeters are generally calibrated to read root-mean-square valves. This also means the ammeter is calibrated to read: a) peak values b) effective values c) average values d) none of these
‘b) effective values ref: p. 5
In the absence of a power-factor meter the power factor of a certain electrical load may be determined by using: a) a wattmeter b) a voltmeter and an ammeter c) a watt-hour meter d) a wattmeter, a voltmeter and an ammeter
d) a wattmeter, a voltmeter and an ammeter ref: p. 22
The junction of the two dissimilar metals of a thermocouple produces a flow of current when the junction is: a) surrounded by a DC magnetic field b) heated c) wet d) all of the above
b) heated ref: p. 13
Which one of the below instrument types would not utilize the effect of current to measure voltage? a) electrostatic b) D’Arsonval c) electromagnetic d) electrodynamometer
a) electrostatic ref: p. 1
The meter measuring mechanism used in recording instruments is basically the same as that used for the conventional indicating instrument except in the area of: a) coil arrangement b) fixed visual scales c) circuit arrangement d) instrument damping setting
d) instrument damping setting ref: p. 16
The circuit measuring arrangement of most ground detector meters shows the neutral points of the ground detector potential transformer primary winding to be: a) fused b) solidly grounded c) connected to its associated meter d) ungrounded
b) solidly grounded ref: p. 20