PFT Equipment Quality Control Flashcards
How often are volumes tested on PFT equipment?
Daily at the minimum
How often are leaks tested on PFT equipment?
Daily at the min
How often is flow linearity tested on PFT equipment?
Weekly at the min
How often is volume linearity tested on PFT equipment?
Quarterly at the min
How often is time adjusted on PFT equipment?
Quarterly at the min
How often is software adjusted on PFT equipment?
Whenever new versions are available
How is volume calibrated on PFT equipment?
Calibration check with 3L syringe.
- must fall within +/- 0.015 L or +/- 0.5% of full scale
- should be 3cmh2o constant pressure for 1 min
How are leaks calibrated on PFT equipment?
3 cmH2O (0.3kpa) constant P for 1 min
How is flow linearity calibrated on PFT equipment?
Test at least 3 different flow ranges (with 3 ml syringe)
- must fall within +/- 0.015 L or +/- 0.5% of full scale
How is volume linearity calibrated on PFT equipment?
1L increments with calibrating syringe measured over entire volume range