LINAC Tolerances Flashcards

1
Q

Daily

X-ray output consistency (all energies)

A

3%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Daily

Laser localization - 3D

A

2mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Daily

Laser localization - IMRT

A

1.5mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Daily

Laser localization - SRS/SBRT

A

1mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Daily

ODI at Iso - 3D

A

2mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Daily

ODI at Iso - IMRT

A

2mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Daily

ODI at Iso - SRS/SBRT

A

2mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Daily

Collimator size - 3D

A

2mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Daily

Collimator size - IMRT

A

2mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Daily

Collimator size - SRS/SBRT

A

1mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Daily

Door Interlock

A

Functional

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Daily

Door closing safety

A

Functional

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Daily

Audio/Visual

A

Functional

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Daily

Radiation area monitor

A

Functional

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Daily

Beam On indicator

A

Functional

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Monthly

X-ray output

A

2%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Monthly

Typical dose rate output accuracy - IMRT and SRS/SBRT

A

2% at dose rate

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

Monthly

Photon bean profile constancy

A

1%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

Monthly

Electron bean profile constancy

A

1%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Monthly

Electron bean energy constancy

A

2% / 2mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Monthly

Light/radiation field

A

2mm or 1% on each side

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Monthly

Light/radiation field - asymmetric

A

1mm or 1% on each side

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Monthly

Gantry/collimator indicators

A

1 degree

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Monthly

Accessory trays

A

2mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Monthly | Jaw position indicators - symmetric
2mm
26
Monthly | Jaw position indicators - asymmetric
1mm
27
Monthly | Cross hair centering
1mm
28
Monthly | Tx couch positioning - 3D & IMRT
2mm / 1 degree
29
Monthly | Tx couch positioning - SRS/SBRT
1mm / 0.5 degree
30
Monthly | Wedge placement accuracy
2mm
31
Monthly | Compensator placement accuracy
1mm
32
Monthly | Latching of wedges & blocking tray
Functional
33
Monthly | Localization lasers - 3D
+- 2mm
34
Monthly | Localization lasers -IMRT
+- 1mm
35
Monthly | Localization lasers - SRS/SBRT
< +- 1mm
36
Monthly | Beam output constancy
2%
37
Monthly | Phase, amplitude beam control
Functional
38
Monthly | In-room respiratory monitoring system
Functional
39
Monthly | Gating Interlock
Functional
40
CT Simulator tolerances Daily CT Number accuracy
CT Number for water - 0 +- 5 HU
41
CT Simulator tolerances Monthly: CT Number accuracy
4 to 5 different materials - 0 +- 5HU
42
CT Simulator tolerances Annually CT Number accuracy
Electron density phantom - 0 +- 5 HU
43
Profile Tolerance
1% from Baseline
44
Calibration TG-51 Tolerance
1%
45
MLC Picket Fence Test Tolerance
Weekly
46
MLC Step/Shoot & Sliding Window
Annual
47
PDD/TMR Tolerance
1%
48
IGRT (KVCT and MVCT) iso check
Daily
49
IGRT (KVCT and MVCT) iso check tolerance
Daily: | 2mm / 1mm (SRS)
50
Dynamic Wedge Output
Monthly
51
Profile Tolerance
1% from Baseline
52
Calibration TG-51 Tolerance
1%
53
MLC Picket Fence Test Tolerance
Weekly
54
MLC Step/Shoot & Sliding Window
Annual
55
PDD/TMR Tolerance
1%
56
IGRT (KVCT and MVCT) iso check
Daily
57
IGRT (KVCT and MVCT) iso check tolerance
2mm / 1mm (SRS)
58
Dynamic Wedge Output
Monthly
59
Annual: | Temporal accuracy of phase/amplitude gating
100ms of expected
60
Annual Calibration of surrogate for respiratory phase/amplitude
100ms of expected
61
Annual | Gating Interlock testing
Functional
62
Annual | X-ray flatness change from baseline
1%
63
Annual | X-ray symmetry change from baseline
+- 1%
64
Annual | Electron flatness change from baseline
1%
65
Annual: | Electron symmetry change from baseline
+- 1%
66
SRS arc rotation mode (range: .5-1.0 MU/deg)
MU set vs. delivered: 1MU or 2% (whichever is greater) Gantry arc set vs. delivered: 1 degree or 2% (whichever is greater)
67
Annual | Calibration TG-51
1% Annual
68
Annual | Output Factors for Small Fields
< 2%
69
Annual | PDD/TMR
1%
70
Annual | TBI PDD X-ray
1%
71
Annual | TSET: electron PDD
1mm
72
Annual | TBI/TSET Output
2%
73
LINAC Wedge Tests
Daily: Functional Monthly: Output Annual: Profile
74
LINAC MLC Tests - Weekly
Picket Fence - visual inspection
75
LINAC MLC Tests - Monthly
Leaf Position Accuracy - 1mm for leaf positions of IMRT field for 4 cardinal gantry angles; Travel speed - loss of leaf speed > 0.5 cm/s; Leaf Setting vs Radiation Field - 2mm
76
LINAC MLC Tests - Weekly
Transmission - 0.5% from baseline; Leaf Position Repeatability - +- 1mm; MLC Spoke shot - < 1mm radius; Rad-Light Coincidence - +-2mm; Step & Shoot test - <0.35cm max error RMS, 95% of error counts<0.35 cm; Moving Window test - <0.35cm max error RMS, 95% of error counts<0.35 cm
77
LINAC Annual Calibration - TG-51:
1. Frequency - Annual 2. Specified - In Water 3. Calibration Factor - NDw 4. Photons - Ion Chamber 5. Photon Measurement depth - 10cm 6 .Photon Calibration depth - Dmax 7. Electrons - Parallel Plate Ion Chamber 8. e' Measurement depth - d reference 9. e' Calibration depth - Dmax 10. Photon measurements use Pb foil - remove e' contamination
78
How often are Ion chambers calibrated?
every 2 years