Procedure for IMRT Commissioning

A
  1. Acceptance Test All equipment
  2. Network Links - IS
  3. Beam Data Collection: PDD, Profiles, Output factors
  4. TPS Beam modeling
  5. Test IMRT dose algorithm
  6. Test Optimization algorithm
  7. MLC Commissioning
  8. R and V System testing
  9. Delivery System
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is the flow for IMRT pt specific QA?

A
  1. MU Check
  2. Absolute Dosimetry (ion chamber - gold standard)
  3. Relative Dosimetry (film - gold standard)
  4. Staff Education
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What does the AAPM allow for total deviation for Rx (absolute dosimetry)?

A

+- 5% (3% is preferred)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What does relative dosimetry measure, what is the tolerance?

A

Evaluation of profiles; 3-5%