Leak Testing Flashcards
A High Pressure Leak Test (method 2)
requires turning the tank on and leaving the flowmeter at __________. Once the manometer reads the pressure in the tank, turn the tank off and wait one hour.
No change = _________________________
Decreased pressure = ____________________.
Leave the flowmeter at Zero
No change = No leak
Decreased Pressure = Leak
What should be performed on your anesthetic machine every day before use
Low pressure leak tests
Seven steps to perform a low pressure leak test
- Assemble machine
- Close pop off valve
- Occlude y piece or universal f circuit
- Turn O2 to 2L/ min or push flush valve
- Fill reservoir bag
- Turn off O2
- Pressure check @ 30cm H2O for 30 sec
You should avoid _______________ inductions to minimize exposure, but if you can’t avoid it, you should have a…
Chamber inductions
Scavenger system
To minimize exposure ET tubes should be used for anesthetic maintenance, avoid ______________ for anesthetic maintenance
Masks
Make sure ET tube cuffs are _________________ and properly _________________ to minimize exposure
Functional
Properly inflated
(if not inflated gas will leak out)
What does the reservoir bag inflating and deflating tell you
There is no leak
To minimize exposure the vaporizer should only be on when ______________, and don’t disconnect until….
Connected
It is turned off
What should you always do before disconnecting a patient to minimize exposure
Empty the reservoir bag into the scavenger
To minimize exposure when a procedure is over, allow the patient to….
Breathe O2