Equipment Failure Flashcards
Equipment failure during anesthesia is uncommon especially since you should be doing a _______________ ______________ every day
Preanesthesia check
One possible problem w equipment failure is CO2 _________________ _________________
Absorber exhaustion
if you have exhaustion of the CO2 absorber, the CO2 in your patient will _______________ and they will exhibit what 5 signs
Increase
1. Tachypnea
2. Tachycardia
3. Arrhythmia
4. Brick red MM
5. Sweat
If your CO2 absorber is exhausted, this will lead to decreased _____________ in ______________ systems
O2
Non rebreathing
Decreased O2 in non rebreathing systems leads to rebreathing and increased….
CO2
6 Possible causes of decreased O2 flow
- Empty tank
- Valves jammed
- Flow meter float lodged
- Disconnected
- Leak
- Incorrect response to problem
What should you do if you have decreased O2 flow (empty tank) with a non rebreather and no negative pressure relief valve
Disconnect
What should you do if you have decreased O2 flow (empty tank) with a rebreathing system and no negative pressure relief valve
Replace the tank
Another possible failure equipment problem is improper assembly. This is why it is important to ____________ the _____________ during your equipment check from the ______________ to the _____________
Follow the circuit
Source to the scavenger
Another possible cause of equipment failure is ET tube…
obstruction