Implant Testing: AC/DP Test before Skin Closure Flashcards

1
Q

Instructor: What is the purpose of the AC/DP testing process before skin closing?

A

Trainee: To confirm the IS is placed correctly and the tongue responds as expected. Adjust placement if needed

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

→ Instructor: What are the recommended stimulation parameters for the procedure?

A

Trainee: The recommended parameters are: frequency at 35Hz, pulse duration at 120 µsec, train length and train interval at 2 seconds each, and amplitude set to 1%.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Instructor: What are the steps for AC/DP testing before skin closure, and what if the response isn’t achieved?

A

Trainee: Partially close the skin, place the AC/DP above the incision, align it with the skin, and check NIM signals and tongue response. Adjust pulse or amplitude as needed. If no response occurs above 20% amplitude, we need to reposition the IS.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly