IFU Exam Questions Flashcards
What is the minimum operating level of the Terumo RS25W on CPB recommended by the company?
200 mls
What ist he max recommended volume that should not be exceeded with the Terumo RX25 oxygenator with an integral hardshell reservoir?
4 L
What is the max gas flow rate that should not be exceeded for the Terumo RX25 Oxygenator?
20 L/min
What is the max flow for the cardiotomy filter in the RX25 oxygenator with hardshell reservoir that should NOT be exceeded?
5 L/min
What is the minimum flow rate for Terumo RX25 oxygenator?
0.5 L/min
Describe what the manufacturer recommends for oxygenator position of a RX25RW oxygenator.
The arterial outflow from the oxygenator is pointing outward toward the field (not toward the perfusionist). with the water line connections facing away from the pump/outside the pump.
What is the minimum blood flow rate recommended by the manufacturer for the Medtronic Affinity NT integrated CVR oxygator is how many L/min?
1 L/min
The max blood flow rate recommended by the manufacturer for the medtronic affinity NT integrated CVR oxygantor is how many L/min?
7 L/min
What is the max water side pressure that should not be exceeded for the MEdtronic Affinity NT integrated CVR is how many pounds per square inch?
30 psi
WHat is hte max cardiotomy flow rate that should not be exceeded for hte medtronic affinity NT integrated CVR is how many L/min?
6
The minimum operating level for the Medtronic Affinity NT integrated CVR is ?
200 mls
The priming volume of the affinity arterial line filter is how many mls?
221 mls
The max flow rate of the affinity arterial line filter is how many L/min?
7 L/min
The pore size of the cardiotomy filter for the Medtornic affinity NT is ?
30 microns
The pore size of hte screen filter in the Myotherm cardioplegia delivery system is ?
150 microns
The blood phase pressure limit of hte myotherm cardioplegia delivery system is ?
750 mmHg
What is the max water side pressure that can be applied to the heat exchanger for the Affinity NT oxygenator?
30PSI
What size are the water points of hte heat exchanger for the Affinity NT Oxygenator?
1/2 ‘’
What kind of filter is the cardiotomy filter for the Affinity NT Oxygenator?
Depth
What material is used to make the heat exchanger bellows of the Myotherm Cardioplegia
Stainless Steel
What type of fitting is used to connect the water lines to the heat exchanger of the Myotherm Cardioplegia Delivery system?
Hansen quick disconnect
What are the Sorin S3 shim sizes (tubing sizes) for 8:1 cardioplegia? (in inches)
Blood: 17/64’’ x 1/16 ‘’
Crystalloid: 3/32’’ x 1/16’’
What are the Soring S3 shim sizes for 1:8 cardioplegia? (in inches)
Blood: 3/32’’ x 1/16’’
Crystalloid: 17/64’’ x 1/16’’
What are the Sorin S3 shim sizes for 2:1 cardioplegia? (in inches)?
Blood: 1/4’’ x 1/16’’
Crystalloid: 3/16’’ x 1/16’’
If you gave 1500 mL of Custodial cardioplegia antegrade in the customary way it is given, how much of the delivery cardioplegia solution is blood/crystalloid?
Blood: 0mL
Crystalloid: 1500 mL