Exam 3 Ch 22- Scatter Production Flashcards
3 main factors that contribute to production of scatter-
kVp, field size, & patient thickness
The only x-ray photons that are incident on an IR are-
those that didn’t interact with the patient
3 main tools for controlling scatter-
beam restriction, grids, & compression (not talked ab a lot)
3 types of x-rays responsible for imaging on a radiograph-
transmitted & scattered (both show up black) & absorbed (white, not talked ab a lot)
image-forming x-rays-
incident on IR
As SID decreases, what increases?
magnification
As kVp increases, what also increases?
-relative compton interaction
-scatter production
As kVp increases, what decreases?
-photoelectrons rapidly
-likelihood of any kind of single photon during a compton interaction
What does compton add to a radiograph?
fog/scatter
compton is scatter, but photons can-
scatter more than once
Why can you not use a low kVp technique?
-increases pt. dose
-would have fewer transmitted photons (black)
collimation reduces ____ ____ & improves ______.
reduces pt. dose & improves contrast
How does collimation help reduce scatter?
by restricting the beam
Field size affects-
amount of scatter produced
As field size increases, what also increases?
scatter
As field size increases, what decreases?
contrast
As patient thickness increases, what also increases?
scatter
with increasing pt. thickness, more x-rays undergo-
multiple scattering
What does pt. thickness result in?
greater average angle of scatter
What can be used to control patient thickness?
sometimes, compression devices
contrast-
visible difference between light & dark areas of an image
contrast resolution-
ability to image & distinguish the difference
as scatter increases, what decreases?
contrast
3 beam restrictors-
aperture diaphragm, cones & cylinders, & variable aperture collimator
Use a grid when part thickness gets above __, ___, or __. or when kVp is ____.
-when part thickness gets above 10, 12, or 13
-kVp is greater than 60
4 types of grids-
parallel, crosses, focused, & moving
2 reasons to use beam restrictors-
-only the tissue being examined should be exposed
-large field size increases scatter & decreases contrast
aperture diaphragm-
simplest & most restrictive beam restricting device (any kind of pitch aperture distorts it)
Aperture diaphragm is designed on film to-
leave 1 cm around field to prove collimation
What kind of construct is the aperture diaphragm?
lead/lead lined metal construct
aperture diaphragm opening is designed to be-
1 cm smaller on all sides of film
cones & cylinders is a modification to-
modification to aperture diaphragm
cones & cylinders resulting exposure-
resulting exposure is circular
cones & cylinders position & distal end determines-
field size
What is 1 problem with cones & cylinders?
cone-cutting
cone-cutting displayed by-
grid cut-off
What is the collimator also known as?
variable aperture diaphragm
collimator 1st stage-
enhances shutters designed to control off-focus radiation
collimator 2nd stage-
shutters work in pairs for square/rectangle fields
collimator allows for
infinite # of field sizes
collimator is more ___ & the least ____.
more versatile & the least __
collimator is the most commonly used-
most commonly used beam restrictor
PBL was mandated by-
US FDA for all x-ray machines in 1974
PBL was removed in 1994 bec-
bucky tray has sensors that detect-
size & alignment of cassette
synchronous motors are actuated to-
adjust to appropriate field size