Ex2 Anesthesia Gas Machine Flashcards
Most fragile part of anesthesia gas machine
Flowtubes
What are flowtubes susceptible to?
Breakage, leakage, inaccuracy d/t dirt, static electricity
Flowmeters - Shape
Variable orifice flowmeter
Tapered - narrow at bottom; wide at top
Annular opening around float is larger at higher flows
Why is oxygen always added last in path of flow?
Chance of producing hypoxic mixture is lessened
Only cylinder safety relief valve that melts at temperature
Fusible plug made of Woods metal
Melts at temperature 157-220 F
Most fragile part of cylinder
Cylinder valve
Only monitor to accurately detect hypoxic mixture
O2 analyzer
O2 Flush valve
Average delivery 50L/min
35-75 L/min
Fail safe is based on
Pressure (Oxygen)
Fail-safe Datex-Ohmeda
Pressure sensor shutoff valve
Flow of all other gases shut off when O2 supply pressure < 20 psi
Drager - Fail safe
Oxygen failure protection device (OFPD)
Reduces supply of other gases proportionally as O2 supply pressure decreases. Shuts off all flow of N2O/air when O2 pressure < 12psi
Second stage regulators are which machine
Datex-Ohmeda
Rod and diagram proportioning system
Drager - oxygen ratio monitor controller
Drager maintains
25% O2
Proportioning systems are based on
FLOW (not pressure)
- which also means concentration
- 25%
Second stage regulators trigger
O2 14 psi
N2O 26 psi
Most dangerous situation regarding hypoxic guard system
Inert gas administration (3rd gas such as Helium) should detect with O2 analyzer
-able to bypass hypoxic guard system w/ Drager
Location of proportioning devices
*ratiometers
Between the O2 and N2O flowmeters
*between fail safe/flowmeters