Electronics Flashcards
Chassis leakage current must be…
less than 300 uA
Amplifier leakage (to the patient) must be…
Less than 10uA
Leakage into the amplifier, from the patient leads, must be…
10uA unattached, or 20uA with patient leads attached
OR grounding must be…
across 1Kohm at 60Hz, less than 20mV
Macroshock
Large deadly current such as “finger in an outlet”
Microshock
Smaller current, but still lethal due to direct application to heart
Ground loops
Occurs when more than one earth ground exists and a voltage difference exists across them
Current can potentially flow from one ground to the other (across the pt) forming a loop
What is a line isolation monitor?
Device that measures combined leakage current to ground from all devices in the OR and alarms if it detects excess current
What is an ammeter?
Measures current
What is an ohmeter?
Measures resistance
What is a voltmeter?
Measures voltage
What is a safety analyzer?
Device used by biomed to perform electrical testing
What do the colors in a power cord mean?
Green=ground
Black=hot
White=neutral
Silver-silver chloride electrodes
surface
Gold electrodes
long time constant, low drift