Decontamination and Disinfection Flashcards

1
Q

Define cleaning

A

Physical removal of microbes (including prions) but not necessarily killing them

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Define disinfection

A
  • Reduction of the number of viable microorganisms but may not inactivate some microbial agents, such as certain viruses and bacterial spores
  • It should make the disinfected object safe to handle
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Define sterilisation

A

Killing and removal of all microorganisms including bacterial spores

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

How should we decontaminate high risk equipment?

A

Sterilisation or single use item

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

How should we decontaminate medium risk equipment?

A
  • Preferably sterilisation
    OR
  • High level disinfection
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

How should we decontaminate low risk equipment?

A

Disinfection

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What does HTM01-05 state about single use instruments?

A

Single use items should never be reused

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What does HTM01-05 state about the number of hand washing and decontamination sinks in decontamination units?

A
  • 1 Hand washing sink

- 2 Instrument decontamination sinks

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What does HTM01-05 state about wrapped instrument storage?

A

Wrapped instruments may be stored up to 1 year

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

List some modes of transmission for microbes (5)

A
  • Direct contact
  • Indirect contact
  • Inhalation
  • Ingestion
  • Inoculation
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What does HTM01-05 state about storage of unwrapped instruments? (2)

Compare storage in clinical vs non clinical areas

A
  • Maximum storage in clinical area is 1 day

- Maximum storage in non-clinical area is 1 week

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What does HTM01-05 require immunisation from?

A
  • HBV

- Tetanus

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Describe a single decontamination room (5)

A
  • Dirty zone
  • Clean zone
  • Separate wash hand basin and storage for PPE in both zones
  • Air flow from clean to dirty
  • Instruments flow from dirty to clean
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What would you find in a dirty zone of a single decontamination room? (5)

A
  • Washing sink
  • Rinsing sink
  • Ultrasonic bath
  • Preliminary inspection area
  • Thermal washer disinfection
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What would you find in a clean zone of a single decontamination room? (4)

A
  • Inspection area
  • Steriliser
  • Package area
  • Storage area
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What are the applications of high risk contamination instruments? (4)

A
  • Penetrate soft tissue
  • Contact bone
  • Enter normally sterile tissues or blood stream
  • Close contact with break in skin or mucous membrane
17
Q

What are the applications of medium risk contamination instruments? (3)

A
  • Contact non-intact skin
  • Mucous membrane
  • Body fluid
18
Q

What are the applications of low risk contamination instruments? (2)

A
  • In contact with intact skin

- Does not come in contact with patient

19
Q

Describe a double decontamination room (5)

A
  • Dirty room
  • Clean room
  • Separate wash hand basin and storage for PPE in both zones
  • Air flow from clean to dirty
  • Instruments flow from dirty to clean
20
Q

What is the aim of cleaning?

What cannot be carried out effectively if instruments are not clean?

A
  • Decrease the bacterial load by thousand fold

- If the instruments are not clean, they cannot be sterilised effectively

21
Q

Name 2 types of automated cleaning?

A
  • Ultrasonic bath

- Thermal washer disinfector

22
Q

Why is it important to avoid letting instruments dry out?

A

Blood is corrosive to stainless steel therefore, do not let instrument dry out

23
Q

What type of detergents are recommended in manual cleaning?

A

Non-foaming detergent

24
Q

What type of brushes are recommended in manual cleaning?

A

Use long handle soft nylon brushes

25
Q

What temperature is recommended in manual cleaning?

A

Water temperature should not be more than 45°C to avoid protein coagulation

26
Q

What type of drying instrument is recommended in manual cleaning?

A

Use disposable lint free cloth (paper towel) to dry instrument

27
Q

What type of instruments can be put in an ultrasonic bath?

A

For metal instrument not plastic (absorb ultrasonic energy)

28
Q

Does an ultrasonic bath disinfect instruments?

A

Only clean but does not disinfect instruments

29
Q

Why is the thermal washer disinfector the preferred cleaning method? (2)

A
  • More efficient in reducing bacterial load during pre- sterilisation cleaning
  • Enhanced staff safety as handling sharp contaminated instruments is significantly reduced
30
Q

Why should Gluteraldehide not be used as a chemical disinfectant?

A

Irritant and sensitise the skin, eyes and respiratory tract

31
Q

Why should Chlorhexidine not be used as a chemical disinfectant?

A

Chlorhexidine binds to protein and makes them more adherent to the instrument

32
Q

What are the recommended chemical disinfectant agents? (2)

A
  • Peracetic acid

- Chlorine releasing agents

33
Q

Sterilisers N cycle;

Air removal?
Suitable for sterilising?
Not suitable for sterilising? (5)

A
  • Non-vacuum
  • Solid
  • Wrapped devices
  • Hollow devices
  • Devices with lumen
  • Heat labile
  • Single use item
34
Q

Sterilisers B cycle;

Air removal?
Suitable for sterilising?
Not suitable for sterilising? (2)

A
  • Vacuum steriliser
  • Most instruments
  • Heat labile
  • Single use item
35
Q

Sterilisers S cycle;

Air removal?
Suitable for sterilising?
Not suitable for sterilising? (2)

A
  • Vacuum steriliser
  • Load specified by manufacturers
  • Heat labile
  • Single use item