CT physics 3 Flashcards
kW of CT
100 kW and above
mA of CT
800mA
focal spot size of CT
0.5-1.2mm2
scan sequences
30s
large heat capacity and fast cooling
up to 8MU, 5MHU/min
what is used to remove low energy photons that dont contribute to image but increase patient dose
flat filter
purpose of beam shaping filter
- dose modification; can attenuate/ enhance frequency of x-ray beam
- scatter reduction; selectively attenuates lower-energy x-rays
- energy selection
- shape modification; match anatomy/ target area
- contrast enhancement
- dose optimization
why is collimation important
- limit patient dose and maximise image quality
- need to collimate slice acquisition in z-axis
- detectors arranged to obtain constant beam width at detector which helps remove scatter
what is the acquired slice thickness determined by
z-axis collimation
- can range from 1 to 160 mm
what does collimator width determine
the width of section sensitivity profile (SSP) which influences spatial resolution z-axis
z-axis (longitudinal axis)
- primary axis od CT scanner
- longitudinal direction of patients body
- corresponds to direction of patient table movement through gantry
what does z-axis determine
the thickness of slices/ sections of body that are imaged
- when patient moved through gantry, creates a series of cross-sectional images that represent different slices of body
- patient translate direction
y-axis (vertical axis)
corresponds to vertical direction as patient positioned on ct scanner table
- represents superior-inferior orientation within body
combination of both axis
defines 3D coordinate system
what does a larger z-axis collimator mean
higher photon flux reaching detectors = less noise and better contrast resolution –> partial volume artifacts