CT Flashcards
CT Artefact - Beam hardening
Beam hardening artefacts appear as streaks and shadows adjacent to areas of high density such as the petrous bone, shoulders, and hips The artefact occurs because the high density anatomy absorbs the lower energy photons while the higher energy photons pass through to the detectors which results in the beam becoming ‘harder’ Iterative reconstruction and bowtie filter reduces beam hardening artefact
CT Artefact - Partial Voluming
Tissues of widely different absorption are encompassed on the same CT voxel producing a beam attenuation proportional to the average value of these tissues Thicker slices and smaller matrix is more prone to partial voluming Mitigated by thinner slices / smaller pixels - however image will appear much noisier unless there is a compensatory increase in the mAs or reduction in pitch
CT Artefact - Ring Artefact
Faulty detector element
CT - Noise
mAs: reduced mAs will decrease the photon flux density and will increase noise kV: decreasing kV will increase contrast, but less photon flux at detector, thereby increasing noise Matrix size / pixel size: reduced pixel size will increase spatial resolution but will decrease photon flux per pixel resulting in increased noise Slice thickness: the thicker the slice, the higher the photon flux in each voxel, reducing noise but reducing spatial resolution Pitch: decreasing the pitch will increase overlapping of slices and increase the photon flux per pixel, reducing noise Patient size: thicker/larger patients will lead to reduced photon penetration (reduced photon flux on the image receptor) leading to increased noise increased chance of scatter in thicker patients reduces the contrast Algorithm: bone algorithm increase noise, but improves spatial resolution (as opposed to soft tissue algorithm which improves contrast)
CT - Spatial resolution
Matrix size: smaller matrix size Slice thickness: thinner slices Sharp algorithm: image appears noisier, but better spatial resolution
CT - Contrast
Ionic markedly hypertonic to achieve sufficient concentration of iodine high osmolality and high viscosity -> 4 fold increase in side effects (vascular, cardiac) vs non-ionic Non-ionic low osmolality Side effects mild = nausea, vomiting, warmth moderate = rash, vasovagal, bronchospasm, laryngeal oedema, vasovagal severe = severe bronchospasm, severe laryngeal oedema, cardiac arrest, anaphylaxis risk of death = 1/170,000 extravasation can result in erythema to tissue necrosis contrast induced nephropathy = acute renal failure within 48 hrs of contrast administration high chance of CIN in at risk patients (eg. diabetes, CKD) therefore patients should be well hydrated Iodinated contrast can cross placenta -> effects on foetus not well documented
CT Fluoroscopy
Display of continuously updated images produced by continuous rotation of a CT tube Generally performed at same kV but lower mA than conventional CT (120kV, 50mA as opposed to 200-300mA) System requirements: Slip rings for continuous scanning Fast tube rotation (1 second) High heat capacity tube to allow for extended scanning times Fast image reconstruction (> 3 images per second) Cine image display allowing last image hold and video recording Bed mounted control / foot switch Rapid image reconstruction: Each image not a complete reconstruction each image uses some new data, but mostly data from previous images eg. tube spins 60 degrees per image, therefore this data added to the other 300 degrees from previous images and the previous 60 degrees is subtracted reconstruction is still via back-projection Usually reconstructed on a 256×256 matrix, rather than the standard 512×512 matrix Dose to patient and operator: Tube current is low to reduce dose but it is concentrated to a smaller area than conventional CT (therefore deterministic effects can occur eg. skin erythema) For patients, the overall dose is comparable to conventional CT For the operator considerable doses can be accumulated, especially to the skin of the hands * Ways to reduce operator dose ensure that the hands are out of the beam when the scanner is operating use of a pair of forceps to manipulate the biopsy or drainage needle from a distance
CT Fluoroscopy Radiation Risk
Risks to the doctor performing the procedure include: effect from radiation exposure, especially on the limb (hand/arm) used to insert the needle/drain within the x-ray beam especially during continuous image acquistion stochastic (cancer induction) and deterministic effects (skin erythema, necrosis) needle stick injury from biopsy/drainage needle Precautions to minimize risks of radiation exposure:* Time: minimise scan times to reduce radiation dose screen intermittently and only when required (as opposed to continuous) make use of previous acquired images to review anatomy (last image hold) Distance: Stand as far away from the x-ray source as possible, though procedural work will require the doctor standing next to the patient the needle/drain etc should be held in needle holders or forceps to avoid having the doctors hands within the x-ray beam * Shielding: wear lead aprons, boots, calf shields, thyroid shield, lead glasses, moveable lead shields, lead gloves if possible Technique: if possible, CT guided procedure can be performed by intermittent scanning planning scan performed marking performed by radiographer on patients skin needle insert, doctor exits the CT room and pt is scanned in region of interest to check position doctor re-enters room and repositions needle etc and exits again to rescan and check position allows the doctor to avoid radiation exposure altogether, though is more time consuming than CT fluoroscopy
Parameters affecting patient dose
- kVp
- mAs
- tube current modulation
- beam shaping filter
- pitch
- scan length
- field of view
- algorithm
Factors affecting patient dose: Patient size
- Higher mAs required for obese patients
Factors affecting patient dose: Rotation Time
- Faster rotation = decreased dose through lower expsosure
- Increased noise
Factors affecting patient dose: Beam Quality
- Higher kVp = more photons produced per mAs = higher dose (by power of 2)
- Higher kVp = decreased subject contrast
Factors affecting patient dose: Scanned Volume
- Decrease the scanned volume to only the clinical area of interest
- Scan patient in one large block rather than two or more smaller blocks
Factors affecting patient dose:Reconstruction algorithm
- Bony algorithm = increased spatial resolution BUT also increased noise
Smoothing algorithm = lower resolution BUT also lower noise
Factors affecting patient dose: Beam Width
- Large collimation reduces the effects of overbeaming as it results in a smaller penumbra
- Over beaming is increased through
- Short beam width (short collimation)
- Large focal spot
- Poor radiographic geometry e.g. short SDD