CLINICAL - Leak Testing Breathing Systems Flashcards
1
Q
Label the components of this breathing system
A
2
Q
Describe how to leak test a breathing system
A
- First connect the breathing system to the anaesthetic machine
- Close the APL valve
- Place your thumb over the patient end of the system
- Press the oxygen flush button to fill the system with oxygen until the reservoir bag appears fully inflated
- Check for leaks via signs of deflation for 10 seonds. You can even give the bag a light squeeze
- Open the APL valve (do not remove your thumb until the APL valve is closed so you do not accidentally leave it closed)
- Remove your thumb
- State if the outer tubing (or just tubing if it is not a coaxial system) and the bag do/do not have a leak
3
Q
Describe how to leak test the inner tube of the coaxial bain system
A
- Turn on the anaesthetic machine
- Set the oxygen flowmeter to 4-6L/min, reading from the top of the bobbin
- Briefly occlude the inner tube at the patient end using a syringe plunger
- If the inner tubing is intact, back pressure will cause the flowmeter bobbin to drop
- Be careful not to be forceful when occluding the inner tube as the force itself can slightly move the bobbin - giving a false reading
- State if the inner tubing does/does not have a leak