humidity of decontamination, general work area, and sterilization equipment room
30 - 60%
humidity of the sterile storage area
should not exceed 70%
decontamination temperature
60ºF - 65ºF
general work area temperature
68ºF - 73ºF
sterile storage area temperature
75ºF or lower
sterilization equipment room temperature
75ºF - 85ºF
air exchanges per hour in the decontamination, general work area, and sterilization equipment room
10
air exchanges per hour in the sterile storage room
4
air pressure in decontamination room
negative
air pressure in the sterile store, general work area, and sterilization equipment room
positive
carts bottom need to be solid and ___-___ inches from the floor
8 - 10
carts and storage shelves should be ___ inches from the wall
2
top of the product on the carts and storage shelves should be ___ inches from the sprinkler head
18
Ethylene Oxide (EtO, EO) destroys microorganisms by means of
Alkylation
Hydrogen Peroxide (H2O2) and Ozone sterilization destroy microorganisms by means of
oxidation