Chart 1: Don, Duff, Handwashing, Cart, and Room Parameters Flashcards

1
Q

humidity of decontamination, general work area, and sterilization equipment room

A

30 - 60%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

humidity of the sterile storage area

A

should not exceed 70%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

decontamination temperature

A

60ºF - 65ºF

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

general work area temperature

A

68ºF - 73ºF

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

sterile storage area temperature

A

75ºF or lower

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

sterilization equipment room temperature

A

75ºF - 85ºF

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

air exchanges per hour in the decontamination, general work area, and sterilization equipment room

A

10

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

air exchanges per hour in the sterile storage room

A

4

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

air pressure in decontamination room

A

negative

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

air pressure in the sterile store, general work area, and sterilization equipment room

A

positive

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

carts bottom need to be solid and ___-___ inches from the floor

A

8 - 10

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

carts and storage shelves should be ___ inches from the wall

A

2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

top of the product on the carts and storage shelves should be ___ inches from the sprinkler head

A

18

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Ethylene Oxide (EtO, EO) destroys microorganisms by means of

A

Alkylation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Hydrogen Peroxide (H2O2) and Ozone sterilization destroy microorganisms by means of

A

oxidation

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q
A
17
Q

the PEL limit of Ethylene Oxide and Hydrogen Peroxide is ___ppm/8 hour time weighted average

A

1

18
Q

the PEL limit of Ozone sterilization is ___ ppm/8 hour time weighted average

A

.1

19
Q
  • has a low boiling point of 51.3ºF
  • 2.5 hour cycle
  • 12-14 hours aeration required
  • uses Bacillus atrophaeus for biological monitoring
A

Ethylene Oxide (EtO)

20
Q
  • uses a hydrogen peroxide solution of 59-95%
  • uses Geobacillus stearothermophilus for biological monitoring
A

Hydrogen Peroxide (H2O2)

21
Q
  • 4.5 hours at temperatures from 85ºF - 94ºF
  • uses Geobacillus stearthermophilus for biological monitoring
A

Ozone (O3)

22
Q

what is the next step after duffing (removing) your gown in decontamination?

A

mask

23
Q

what is the final step donning on (putting on) PPE in decontamination?

A

gloves