Characteristics of Clinical Beams LT2 Flashcards
Define Beam Size
FWHM @ 10cm deep
Define penumbra
Distance between 80% and 20% dose levels @ 10cm deep
What happens to beam profile with increasing depth?
- beam widens (divergence)
- penumbra widens (more scatter)
- Dose decreases
- Flattening filter makes it flattest at 10cm
- Horny at lower depths
- Rounded & concave at deeper parts
How is wedge angle defined?
Angle between isodose line and normal to central axis at 10cm deep
Three causes of penumbra
- Geometric (extended source)
- Transmission (through collimator)
- Dosimetric (scatter in patient)
What does geometric penumbra depend on?
- Width of extended source
- Collimator position
- SSD
Penumbra at depth d = (s(SSD + d) -SSD) / SSD
s=width of source
d=depth
What affects transmission penumbra?
Energy of beam
How is dosimetric penumbera affected by energy?
Lower energy = more lateral scatter hence wider penumbera
Which types of penumbra dominate at high energies?
Transmission & Geometric
How do you define a field size factor?
Why do you need them?
What two parts is it made up of?
Ratio of dose at 10cm depth under reference conditions to different collimator condition
Beam size increases, scatter increases, dose increases
FSF is made up of phantom scatter factor and head scatter factor
Head Scatter Factor
What does it depend on?
How do you measure?
What component causes it?
- Difficult to quantify
- Depends on make of LINAC & is fn of field size
- Measure by keeping irradiated volume constant (removing phantom scatter)
- Mostly from flattening filter (3-4% dose to patient)
*
Phantom Scatter Factor (SP)
What does it depend on?
How do we measure?
- Depends on beam quality
- Difficult to measure, infer from Sc (head scatter) and ST (total scatter)
- Mostly from flattening filter
Monitor Units:
- Definition
- Why do we have them?
- Amplifier adjusted until:
- 1MU = 1Gy under calibration conditions
- (e.g. 1MU = 1Gy at dmax for a 10x10 field)
- Have them to stop the LINAC when required number of MUs is given from TPS. LINAC calibration defines reference Dose/MU for dosimetry
When calculating Dose/MU to a patient, what deviations from reference conditions might you need to think about? (x6)
- Depth
- Treatment distance
- Collimator setting
- Shape/size of area?
- Attenuators e.g. wedges
- patient heterogeneities
What equation system should you use for:
Isocentrix
fixed SSD
- Isocentric - TPR (i.e. you are always prescribing dose to the same point, but you might adhd more or less water on top hence TPR)
- Fixed SSD - PDD (i.e. you are always prescribing dose to the surface so you only really care about how the dose drops off with depth