Chapter 8 - HPLC maintenance and troubleshooting Flashcards

1
Q

Frequently encountered troubleshooting categories (4)

A
  1. pressure
  2. baseline
  3. peak
  4. data
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

The typical wearable item in the autosampler is the

A

injector rotor seal.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

A rotor seal can last approximately how many injections?

A

30,000

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Verify that a problem exists by

A

repeating the experiment.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Baseline drift in a UV detector often reflects

A

the change of the energy output of the UV lamp with time.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Spikes on the baseline are likely caused by

A

air bubbles out-gassing in the detector flow cell.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Peak problems such as poor resolution, broad, split, tailing, or fronting peaks, and extra peaks are most often caused by

A

the column.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Fronting peaks are caused by

A

column overload.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

OOS stands for

A

Out of specification

How well did you know this?
1
Not at all
2
3
4
5
Perfectly