Chapter 8 Flashcards
Pacemaker
Battery powered device that delivers an electrical current to the heart to stimulate depolarization
Consists of a pulse generator and pacing leads
Pacing lead
Insulated wire that is used to carry an electrical impulse from the pacemaker to the patients heart and information about the hearts electrical activity back to the pacemaker
Triggering - start
Inhibition - stop
Pulse generator (power source)
Houses a battery and electronic circuitry
Temporary pacing routes
An artificial pacemaker can be external (temporary intervention) or implanted
Pulse generator of temporary pacemaker is located externally
External pacemakers may be used to control transient disturbances in the heart rate or conduction
Can be accomplished through transvenous, epicardial, or transcutaneous means
Transvenous pacing
Transvenous pacemakers stimulate the endocardium of the right atrium of ventricle (or both)
Electrode is introduced into a central vein, such as the subclavian, femoral, bronchial, internal jugular, or external jugular vein
Epicardial pacing
Epicardial pacing is the placement of pacing leads directly onto or through the epicardium
May be used when a patient is undergoing cardiac surgery and the hearts outer surface is easy to reach
Transcutaneous pacing (TCP)
Uses electrical stimulation through two pacing pads positioned on a patients torso to stimulate contraction of the heart
Electrical stimulus exists from the negative terminal on the machine (and subsequently the negative electrode) and passes through the chest wall to the heart
Range of output current varies depending on the manufacturer
(Doesn’t work too well)
Transcutaneous pacing - complications
-coughing
-skin burns
-interference with sensing
-discomfort from electrical stimulation of the skin and muscles
-failure to recognize that the pacemaker is not capturing
-tissue damage, including third degree burns
-pacing threshold changes
Permanent pacemakers
Patients who have chronic dysrhythmias that are unresponsive to medication therapy and that result in decreased cardiac output may require the surgical implantation of a permanent pacemaker or an implantable cardioverterdefibrillator (ICD)
Pacemakers and ICDs are called cardiovascular implantable electronic devices (CIEDs)
Used to treat disorders of the Sinoatrial (SA) node, disorders of the AV conduction pathways, or both, that produce signs and symptoms as a result of inadequate cardiac output
Pacemakers pulse generator is usually implanted under the local anesthesia into the subcutaneous tissue of the anterior chest just below the right or left clavicle
Pacemaker leads
Pacemaker lead systems may consist of single, double, or multiple leads
-separate lead is used for each heart chamber paced
Electrode
- the exposed portion of the pacing lead that is placed in direct contact with the heart
Unipolar electrode
-has one pacing electrode located at its distal tip
-less commonly used than bipolar lead systems
Bipolar lead system
-contains a positive and negative electrode at the distal tip of the pacing lead wire
Leadless pacemakers
Consists of self contained generator and lead systems
Inserted through the femoral vein and into the right ventricle, eliminating the need for a chest incision and creating a pocket in the patients chest to house a pulse generator
First generation leadless pacemakers provide only single chamber ventricular pacing
Pacing principles
Pacing occurs when the pacemakers pulse generator delivers energy (milliamperes [mA]) through the pacing electrode to the myocardium
A pacemaker responds to the information received either by sending a pacing stimulus to the heart (triggering) or by not sending a pacing stimulus to the heart (inhibiting)
Fixed rate pacemakers
Continuously discharge at a preset rate (usually 70-80 impulses/min) regardless of patients heart rate or metabolic demands
Simple circuitry
Does not sense patients own cardiac rhythm
-may result in competition between the patients cardiac rhythm and that of the pacemaker
Not often used today
Demand pacemakers
Discharge when the patients heart rate drops below the pacemakers rate limit
Expressed in paced pulses per minute (PPM)
Set up depending on what your heart needs
Capture
Capture
-the successful conduction of an artificial pacemakers impulse through the myocardium, resulting in depolarization
Electrical capture
-can be seen as a pacemaker spike followed by an atrial or ventricular complex on the ECG depending on the chamber that is being paced
Mechanical capture
-assessed by palpating the patients pulse
Atrial pacing
Pacing electrode placed in the right atrium
Produces a carmaker spike followed by a P wave
May be used when SA node is diseased or damaged but conduction through AV junction and ventricles is normal
Ventricular pacing
Pacing electrode placed in right ventricle
Produces pacemaker spike followed by a wide QRS, resembling a ventricular ectopic beat
Pacemaker spike in front of QRS (wide)
Pacemaker syndrome
Because a ventricular demand pacemaker does not coordinate pacing with the patients intrinsic atrial rate, it can result in asynchronous contraction in the atrium and ventricle
Loss of AV synchrony can result in a loss of the atrial contribution to cardiac output, decreased stroke volume and decreased cardiac output
Dual chamber pacemaker
Two lead system placed in the heart
-one lead is placed in the right atrium
-a second lead is placed in the right ventricle