Chapter 7 Quiz Flashcards

1
Q

All of the following are components of a fluoro system EXCEPT:

a.) video monitoring system
b.) image plate reader
c.) xray tube
d.) generator
e.) image intensifier

A

b.) image plate reader

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Reproducibility for fluoro equipment should be within:

a.) 5%
b.) 10%
c.) 3%
d.) 2%

A

a.) 5%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Performance testing should be performed every ____________.

a.) month
b.) week
c.) 6 months
d.) year

A

c.) 6 months

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Making a 30 second exposure at maximum kvp and ma is testing the:

a.) reciprocity
b.) ma accuracy
c.) kvp accuracy
d.) heat sensor

A

d.) heat sensor

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

The maximum ESE for fluoro is:

a.) 50 R/min
b.) 10 R/min
c.) 150 R/min
d.) 100 R/min

A

b.) 10 R/min

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Which of the following is not a “tube” type of television camera?

a.) CCD
b.) orthicon
c.) plumbicon
d.) vidocon

A

a.) CCD

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Which part of an image intensifier is made of cesium iodide?

a.) output phosphor
b.) input phosphor
c.) rotating anode
d.) photocathode

A

b.) input phosphor

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What are the three types of electric display devices?

a.) LCD
b.) TFT
c.) CRT
d.) plasma

A

a.) LCD
c.) CRT
d.) plasma

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

A motion picture camera is called:

a.) DVR
b.) spot film camera
c.) cinefluorography
d.) spot film device

A

c.) cinefluorography

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Which part of the image intensifier absorbss light and emits electrons?

a.) input phosphor
b.) photocathode
c.) output phosphor
d.) rotating anode

A

b.) photocathode

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

The ability to resolve relatively large objects with similar density is called:

a.) low contrast resolution
b.) minification
c.) flux gaIn
d.) high contrast resolution

A

a.) low contrast resolution

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Which part of an image intensifier is made of zinc and cadmium?

a.) rotating anode
b.) input phosphor
c.) photocathode
d.) output phosphor

A

d.) output phosphor

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Fluoro equipment should always have the following inspections:

a.) performance
b.) enviromental
c.) digital
d.) visual

A

a.) performance
b.) enviromental
d.) visual

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

An artifact caused by magnetic field is called:

a.) S distortion
b.) vignetting
c.) barrel disortion
d.) pincushion distortion

A

a.) S distortion

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

In digital fluoroscopy, the _____________ adjusts radiation output for changes in part thickness.

a.) HVL
b.) PBL
c.) AEC
d.) AERC

A

d.) AERC

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

_____________ is useful in myelograms and arthrograms when an image is necessary to be on the display monitor but the xray beam is off.

a.) image restoration
b.) last image hold
c.) road mapping
d.) filtering

A

b.) last image hold

17
Q

In digital fluoroscopy which part of the flat panel detector absorbs xrays and emits light?

a.) none of these
b.) input phosphor
c.) output phosphor
d.) rotating anode

A

a.) none of these

18
Q

Which part of an image intensifier is made of cesium and antimony?

a.) input phosphor
b.) photocathode
c.) output phosphor
d.) rotating anode

A

b.) photocathode

19
Q

What are the two ways that the output phosphor and the television camera can be linked?
1.) CCD
2.) orthicon
3.) lens coupling
4.) fiber optics
5.) closed circuit TV

a.) 3 and 4
b.) 1 and 2
c.) 2 and 5
d.) 1 and 5

A

a.) 3 and 4

20
Q

A decrease in brightness at the lateral edges of the image is called:

a.) vignetting
b.) veiling glare
c.) S distortion
d.) barrel distortion

A

a.) vignetting

21
Q

For a non DR system, the most common type of image intensifier is the:

a.) tube type
b.) flat panel

A

a.) tube type

22
Q

A heat sensor alarm should activate when ______% of the heat capicity is acheived.

A

75%

23
Q

Which of the following is not part of the image intensifier tube?

a.) anode
b.) filament
c.) output phosphor
d.) photocathode

A

b.) filament

24
Q

Which mode results in higher patient dose?

a.) flux gain
b.) magnification
c.) minification

A

b.) magnification