Battery Life, LEDs, and Alarms (oh my!) Flashcards
Naída CI Q70 Power Operating Time
(for use with HiRes Optima)
BATTERY TYPE- 2 Zinc-Air Cartridge
AVERAGE (HRS)- 31
RANGE (HRS)- 56
BATTERY TYPE- 110 PowerCel
AVERAGE (HRS)- 12
RANGE (HRS)- 17
BATTERY TYPE- 170 PowerCel with FM
AVERAGE (HRS)- 12
RANGE (HRS)- 16
BATTERY TYPE- 170 PowerCel w/o FM
AVERAGE (HRS)- 18
RANGE (HRS)- 27
BATTERY TYPE- 230 PowerCel
AVERAGE (HRS)- 25
RANGE (HRS)- 36
BATTERY TYPE- AAA PowerPak
AVERAGE (HRS)- 129
RANGE (HRS)- 183
Naída CI Q70 PowerCel Charger LEDs
COLOR- Blue
INDICATION- PowerCel charging
COLOR- Green
INDICATION- PowerCel fully charged
COLOR- Red
INDICATION- Fault condition
COLOR- No light
INDICATION-
- PowerCel fully charged (confirm by waiting 1 minute for LED to turn green)
- Charging zinc-battery charger attached
- Potential fault with PowerCel (if LED doesn’t turn green after 1 minute, should be considered faulty and replaced)
Naída CI Q70 LED Functionality-
Orange
BEHAVIOR- Blinks at start-up
PROGRAMMABLE- No
INDICATION- • 4 quick blinks: battery is fully charged
- 2-3 blinks: battery is sufficiently charged ot power the Naída CI
- 1 quick blink: battery is nearly depleted
- No blinking: depleted battery
BEHAVIOR- Solid
PROGRAMMABLE- Yes
INDICATION- The battery is almost depleted.
BEHAVIOR- Blinks twice every 3 seconds
PROGRAMMABLE- Yes
INDICATION- Battery is almost depleted and cannot support stimulation (Standby Mode)
BEHAVIOR- Fades out
PROGRAMMABLE- No
INDICATION- Naída CI is entering Standby Mode
Naída CI Q70 LED Functionality-
Red
BEHAVIOR- Blinks once per second
PROGRAMMABLE- Yes
INDICATION- Loss of lock with implant.
BEHAVIOR- Blinks rapidly (more than once per second)
PROGRAMMABLE- No
INDICATION- IntelliLink is enabled and Naída CI is connected to the wrong implant.
BEHAVIOR- Solid
PROGRAMMABLE- No
INDICATION- Sound Processor Error Condition. Fully remove and re-insert battery to reset processor.
BEHAVIOR- Blinks 5 times
PROGRAMMABLE- No
INDICATION- Response to AB myPilot request to “Find Paired Devices”. The Right paired device will idenitfy itself with this LED pattern.
Naída CI Q70 LED Functionality-
Green
BEHAVIOR- Flickers in response to loud input
PROGRAMMABLE- Yes
INDICATION- The sound processor and microphone are responding to sound.
BEHAVIOR- Blinks at start-up, after battery status, and upon program change
PROGRAMMABLE- No
INDICATION- 1-5 blinks indicating current Program
BEHAVIOR- Solid
PROGRAMMABLE- No
INDICATION- A processor that is not yet programmed.
BEHAVIOR- Blinks 4 times
PROGRAMMABLE- No
INDICATION- Response to the AB myPilot request to “Find Paired Devices”. The Left paired device will identify itself with this LED pattern.
Naída CI Q70 Internal Alarms
BEHAVIOR- Beeps upon program change
PROGRAMMABLE- Yes
INDICATION- 1-5 beeps indicating current Program
BEHAVIOR- Short Beep upon increase/decrease in volume
PROGRAMMABLE- Yes
INDICATION- Beeps once per press of volume control either up or down. A double beep will be heard when the following settings are reached: top of volume range; baseline setting; bottom of volume range.
BEHAVIOR- Long Beep (once every 15 minutes)
PROGRAMMABLE- Yes
INDICATION- Low Battery
Neptune LED Status & Functionality
Orange
programmable feature, enabled by default
STATUS- Battery Status
BEHAVIOR- Blinks at start-up
INDICATION- 1-4 quick blinks indicates level of charge with 4 meaning fully-charged battery
BEHAVIOR- Solid
INDICATION- Battery almost depleted
BEHAVIOR- Blinks 2x every 2 seconds
INDICATION- Battery depleted and cannot support stimulation
BEHAVIOR- Fades out
INDICATION- Powering down
Neptune LED Functionality-
Red
programmable feature, enabled by default
STATUS- System Status
BEHAVIOR- Blinks 1x per second
INDICATION- Loss of lock
BEHAVIOR- Blinks rapidly
INDICATION- Wrong implant
BEHAVIOR- Solid
INDICATION- Error Condition
Neptune LED Functionality-
Green
programmable feature, enabled by default
STATUS- Microphone and Program Status
BEHAVIOR- Flickers in response to loud inputs
INDICATION- Mic and sound processor are responding to sound
BEHAVIOR- Blinks at start-up
INDICATION- Program position
BEHAVIOR- Solid
INDICATION- Empty program slot or monitor position
Neptune Audible Alarms
Provide info about system status and battery life
programmable feature, disabled by default
BEHAVIOR- Beeps once per second
SOFTWARE MANAGEMENT- Programmable
INDICATION- Loss of Lock with implant
BEHAVIOR- Beeps rapidly
SOFTWARE MANAGEMENT- Programmabale
INDICATION- Wrong implant connected
BEHAVIOR- Beeps slowly, once every 5 seconds
SOFTWARE MANAGEMENT- Programmable
INDICATION- Low battery
BEHAVIOR- 2 beeps every 3 seconds
SOFTWARE MANAGEMENT- Programmable
INDICATION- Battery is depleted and cannot support stimulation
BEHAVIOR- Continuous beep
SOFTWARE MANAGEMENT- Not Programmable
INDICATION- Sound processor error condition
Neptune Power Options
BATTERY TYPE- Lithium AAA battery
AVERAGE (HRS)- 20
RANGE (HRS)- 11-30
BATTERY TYPE- Alkaline AAA battery
AVERAGE (HRS)- 12
RANGE (HRS)- 6-21
BATTERY TYPE- Rechargeable NiMH AAA battery
AVERAGE (HRS)- 13
RANGE (HRS)- 8-21
Neptune Connect Sync Light
BEHAVIOR- One green flash
INDICATION- Neptune Connect and processor settings are synced. Any changes will take effect on the processor.
BEHAVIOR- Flashes red once per second
INDICATION- A mismatch exists between Neptune processor settings and Neptune Connect settings. In order to sync, turn volume dial all the way down (counter-clockwise); verify green light flashes once; then adjust the controls to desired position(s).
BEHAVIOR- Solid red light
INDICATION- Error condition. If this occurs, remove and reattach Neptune Connect.