AIM: Ch 10: CT Flashcards
It is the moving part of the scanner apparatus
Gantry
The data that were collected in this scanning pattern correspond to attenuation measurements that are parallel to each other, and this geometry is called a ____. This is a more general geometrical concept, and it has utility in some nuclear and x-ray tomography geometries till today
Parallel beam projection
In the first years of CT, the x-ray tube and detectors used a linear scanning trajectory, as illustrated in Figure 10-8.
It refers to the data collected at a specific angle of interrogation of the object, and this term is synonymous with the terms profile or view
Projection
Rays are individual attenuation measurements that correspond to a line through the object defined at one end by the x-ray source and at the other end by a detector.
A projection is a collection of rays.
The individual rays in this geometry each correspond to a line integral that spans between the x-ray source and the x-ray detector; however, there are a lot more detectors in this geometry compared to the parallel beam geometry
True fan beam geometry
For typical 64 to 128 detector array CT scanners, the fan angle is approximately ____ degrees, while the full cone angle is about ____ degrees
60
2.4
Figure 10-10 shows a more accurate depiction of a modern multislice CT scanner, where the fan angle defines the “fan beam geometry,” but there is nonnegligible x-ray beam divergence in the orthogonal direction, which gives rise to the concept of narrow cone beam geometry.
Some niche CT scanners make use of flat panel detectors for the detector array. This geometry represents a ____, where the cone angle is almost as great as the fan angle. The use of the planar flat panel detector system leads to a 2D bank of detectors, which has no curvature, and so correction methods are needed for the different distances from source to detector, giving rise to inverse square law and heel effect–based differences in fluence to each detector element in the 2D array.
Full cone beam geometry
Flat panel detectors are used for cone beam systems in dental and maxillofacial CT systems, breast CT systems, and in radiation therapy imaging applications where the cone beam CT system is mounted orthogonally to the high-energy linear accelerator treatment beam.
The ____ is the center of rotation of the CT gantry, and in most cases (but not all), it is also the center of the reconstructed CT image—that is, pixel (256, 256) on the 512 x 512 reconstructed CT image
Isocenter
The ____ is defined by the physical extent of the curved detector arrays (the fan angle)
Maximum FOV
It is defined by the ratio of B and A where A is the source-to-isocenter distance and B is the source-to-detector distance. This factor helps in defining the actual width of the detector by getting the product of this factor and the minimum CT slice thickness.
Magnification factor, M
Almost all modern CT scanners have a design where the x-ray tube and detector are attached rigidly to the rotating gantry, and this leads to a so-called third-generation or ____
“Rotate-rotate” geometry
Thus, as the gantry rotates, the x-ray tube and detector arrays stay in a rigid alignment with each other. This fixed geometry allows the use of an antiscatter grid in the detector array.
The grid septa are aligned with the dead space between individual detectors, in an effort to preserve the geometrical detection efficiency of the system.
As the width of the x-ray beam has increased with multidetector array CT (to 40, 80 mm, and larger), there is greater need for more aggressive scatter suppression. At least one manufacturer uses a 2D grid on its large multidetector array CT system.
The x-ray tube is mounted onto the gantry such that the plane of the anode disk is parallel to the plane of gantry rotation (Fig. 10-14A), which is necessary to reduce ____ that would add significant torque to the rotating anode if mounted otherwise.
Gyroscopic effects
Furthermore, this configuration means that the anode-cathode axis and thus the heel effect run parallel to the z-axis of the scanner. This eliminates heel effect–induced spectral changes along the fan angle, which is important. The angular x-ray output from an x-ray tube can be very wide in the dimension parallel to the anode disk but is quite limited in the anode-cathode dimension (Fig. 10-15), and so this x-ray tube orientation is necessary given the approximately 60-degree fan beam of current scanners.
CT gantry rotation speeds are approaching ____ (rotation/s), and these high angular velocities create enormous g-forces on the components that rotate
5 rotations per second (0.20 s rotation time)
While the x-ray detectors are moving in unison with the x-ray source, the patient is stationary, and thus the 1.5-mm circumferential displacement of the x-ray source over the time it takes to acquire one projection can lead to THIS ARTIFACT and a loss of spatial resolution. To compensate for this, some scanners use a magnetic steering system for the electrons as they leave the cathode and strike the anode.
Motion artifacts
One CT manufacturer also uses a ____ to provide oversampling in the z-dimension of the scan. This x-ray tube design combines high power with the ability to stagger the position of the x-ray focal spot inside the x-ray tube, as illustrated in Figure 10-17A. The electron beam is steered using magnetic fields provided by steering coils, and due to the steep anode angle, modulating the electron beam landing location on the focal track causes an apparent shift in the source position along the z-axis of the scanner.
Focal spot steering approach
Another manufacturer uses a different magnetic electron beam steering approach (Figure 10-17B) to achieve a similar effect.
For helical (spiral) scans, this approach leads to an oversampling in the z-dimension, which leads to Nyquist—appropriate sampling in z
Lowest interaction probability in the range of effective x-ray energies in typical CT spectra overlaid on the mass attenuation coefficients for soft tissue
Rayleigh scattering
Highest interaction probability in the range of effective x-ray energies in typical CT spectra overlaid on the mass attenuation coefficients for soft tissue
Compton scattering
At all x-ray tube voltages, give the HU for air and water
-1,000 HU for air
0 for water
When referring to the location in the body, the term ____ is used
Volume element (voxel)
A beam shaping filter, also called a ____, is used on all whole body CT scanners and is located in the x-ray tube assembly. The shape of the filter is designed to attenuate more toward the periphery of the field, which tends to make the signal levels at the detector more homogeneous.
Bow tie filter
The standard adult head is about 17 cm in diameter, whereas the standard adult torso ranges from 24 up to 45 cm or greater, with an average diameter ranging between 26 and 30 cm. Because of these large differences, all commercial CT scanners make use of a minimum of two bow tie filters—a head and a body bow tie. The head bow tie filter is also used in pediatric body imaging on most scanners.
T/F: The bow tie filter has consequences both on patient dose and on image quality.
True
The bow tie filter is a very useful tool because it reduces patient dose with no loss of image quality
NOISE - from central rays which have lower x-ray fluence and higher statistical variance vs peripheral rays which have higher x-ray fluence and lower statistical variance (lower noise)
To improve the detection efficiency of the scintillator material for CT imaging, the scintillation crystals (Gd2S2O and other materials as well) are ____ to increase physical density and light output. It involves heating up the phosphor crystals to just below their melting point for relatively long periods of time (h), and there are many other details to the process. In the end, densification occurs, and the initial scintillating powder is converted into a high-density ceramic
Sintered
They provide power to the detector array and receive the electronic signals from each photodiode
Electronic modules
The electronics module has gain channels for each detector in the module and also contains the analog-to-digital converter, which converts the amplified electronic signal to a digital number
In this scanner, the detectors were 13 to 16 mm wide, and the x-ray beam collimation was used to determine the slice thickness. Collimator settings ranged from 1 mm (thin slice) to 10 mm (thick slice) or larger. This integrated the x-ray signal over the entire width of each detector, and so changing slice thickness was solely a function of the collimation
Single detector array scanner
In this scanner, the slice thickness and x-ray beam width are decoupled. The slice thickness is determined by the detector configuration, and the x-ray beam width is determined by the collimator.
MDCT
Because the beam shape is so different in the penumbra region compared to the center of the beam, the penumbra region is to be avoided, and hence the penumbra is positioned outside of the active detector arrays—those x-rays essentially strike lead shielding on the sides of the detector assembly or inactive detectors. This is called ____.
Overbeaming
The consequence of overbeaming is that for MDCT scanners, the geometrical efficiency is less than the 100% geometrical efficiency of a single detector array scanner.
Once the patient is on the table and the table is moved into the gantry bore, the technologist performs a preliminary scan called the ____
CT radiograph
This image is also called the scout view, topogram, scanogram, or localizer; however, some of these terms are copyrighted to specific vendors.
The PA CT radiograph is preferred over the AP to reduce breast dose in women and girls.
It is at this point in the scanning procedure that all the CT scan parameters are set, usually using preset protocols. For a basic CT scan, these parameters include kV, mA, gantry rotation time(s), type of scan (helical or axial), direction of scan, pitch, detector configuration, reconstruction kernels(s), mA modulation parameters, if used, and so on
It is the basic step-and-shoot mode of a CT scanner
Axial (also called sequential) CT scan
With the advent of ____, it is common to acquire contiguous CT images during an axial acquisition. This implies that between each axial acquisition, the table moves a distance D, which is essentially equal to the width of the detector array at isocenter (nT). This results in a series of images in the CT volume data set, which are contiguous and evenly spaced along the z-direction.
MDCT
In practice, the x-ray beam width is slightly wider than the distance nT, and so the series of axial scans results in some x-ray beam overlap between the axial acquisitions (Fig. 10-33)