Workstation 1 Flashcards

1
Q

What is the normal range for the oxygen pressure on the anesthesia workstation?

A

Normal range is 40-60 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What is the normal range for the nitrous oxide pressure on the anesthesia workstation?

A

Normal range is 40-55 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What is the normal range for the air pressure on the anesthesia workstation?

A

Normal range is 30-50 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

True or False: The pipeline pressure is typically higher than the cylinder pressure.

A

True.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What is the pressure relief valve set at for the anesthesia workstation?

A

The pressure relief valve is set at 55 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What should the pressure be at the common gas outlet of the anesthesia workstation?

A

The pressure should be 25-55 psi at the common gas outlet.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What is the minimum pressure required at the common gas outlet for the ventilator to work properly?

A

The minimum pressure required is 30 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

What is the minimum pressure required at the common gas outlet for the vaporizers to work properly?

A

The minimum pressure required is 25 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

What is the normal range for the oxygen pressure in the oxygen flush valve?

A

Normal range is 50-70 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What is the normal range for the oxygen pressure in the oxygen pressure failure device?

A

Normal range is 20-30 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What is the pressure needed for the oxygen pressure failure device to activate?

A

The pressure needed is below 20 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

What is the pressure needed for the oxygen pressure failure device to reset?

A

The pressure needed is above 30 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

What is the pressure needed for the oxygen pressure failure device to alarm?

A

The pressure needed is below 20 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

What is the pressure needed for the oxygen pressure failure device to silence the alarm?

A

The pressure needed is above 30 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

What is the normal range for the oxygen pressure in the hypoxic guard system?

A

Normal range is 20-30 psi.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

What is the pressure needed for the hypoxic guard system to alarm?

A

The pressure needed is below 20 psi.

17
Q

What is the pressure needed for the hypoxic guard system to silence the alarm?

A

The pressure needed is above 30 psi.

18
Q

What is the normal range for the oxygen pressure in the oxygen supply failure alarm?

A

Normal range is 20-30 psi.

19
Q

What is the pressure needed for the oxygen supply failure alarm to activate?

A

The pressure needed is below 20 psi.

20
Q

What is the pressure needed for the oxygen supply failure alarm to reset?

A

The pressure needed is above 30 psi.

21
Q

What is the pressure needed for the oxygen supply failure alarm to silence the alarm?

A

The pressure needed is above 30 psi.

22
Q

What is the normal range for the nitrous oxide pressure in the nitrous oxide pressure failure alarm?

A

Normal range is 20-30 psi.

23
Q

What is the pressure needed for the nitrous oxide pressure failure alarm to activate?

A

The pressure needed is below 20 psi.

24
Q

What is the pressure needed for the nitrous oxide pressure failure alarm to reset?

A

The pressure needed is above 30 psi.