Workstation 1 Flashcards
What is the normal range for the oxygen pressure on the anesthesia workstation?
Normal range is 40-60 psi.
What is the normal range for the nitrous oxide pressure on the anesthesia workstation?
Normal range is 40-55 psi.
What is the normal range for the air pressure on the anesthesia workstation?
Normal range is 30-50 psi.
True or False: The pipeline pressure is typically higher than the cylinder pressure.
True.
What is the pressure relief valve set at for the anesthesia workstation?
The pressure relief valve is set at 55 psi.
What should the pressure be at the common gas outlet of the anesthesia workstation?
The pressure should be 25-55 psi at the common gas outlet.
What is the minimum pressure required at the common gas outlet for the ventilator to work properly?
The minimum pressure required is 30 psi.
What is the minimum pressure required at the common gas outlet for the vaporizers to work properly?
The minimum pressure required is 25 psi.
What is the normal range for the oxygen pressure in the oxygen flush valve?
Normal range is 50-70 psi.
What is the normal range for the oxygen pressure in the oxygen pressure failure device?
Normal range is 20-30 psi.
What is the pressure needed for the oxygen pressure failure device to activate?
The pressure needed is below 20 psi.
What is the pressure needed for the oxygen pressure failure device to reset?
The pressure needed is above 30 psi.
What is the pressure needed for the oxygen pressure failure device to alarm?
The pressure needed is below 20 psi.
What is the pressure needed for the oxygen pressure failure device to silence the alarm?
The pressure needed is above 30 psi.
What is the normal range for the oxygen pressure in the hypoxic guard system?
Normal range is 20-30 psi.