What parameters or 102 lead test performed at?
1.0 mA 20 Hz 500 µs 30 seconds on 60 seconds off
What does IFI stand for?
And what does yellow mean on different generators?
Intensified follow-up indicator
8 to 18% battery on 103 and 104 generators
5 to 11% battery on 105 generators
What are the parameters of the system diagnostic test?
1.0 mA
20 Hz
500 µs and
30 seconds on
What does N EOS mean? And by generator type?
Near end of service
0 to 8% battery on the 103 and 104 generators
0 to 5% battery on 105 generators
What are the 8 setting parameters?
Output current in mA Signal frequency in Hz Pulse width in microseconds Signal on time signal off time MAGNET: Output current in mA Signal frequency in Hz Pulse width in microseconds
What are clinical concerns of potential end of service?
Increase and seizure activity
Painful or Erratic stimulation
Report of decreased perception of stimulation or no stim when using magnet
What are MRI conditions that are safe?
Implant set to 0.0 mA
Transmit and receive only
Coil type is 1.5 to 3.0 Tesla in a static magnetic field
Describe testing impedances on 102 batteries
Make sure patient can tolerate 1.0 mA
System diagnostics test assesses lead impedance and integrity
Normal mode assess his generator to deliver programmed output
How can you tell high impedances on pulse generators? 102 or older
DCDC is 4-7 which coincides with high impedance
Accessory pack part number and contents
Tiedowns
hex screwdriver and
2 x test resistors - one single pin and one dual pin
What color is sterilization indicator went good? And when bad?
Good = gold to bronze
Bad = Red or black
What is the part number of the programming wand?
201???? Need to verify this
Is the tunneler compatible with all lead types? And what is the part number?
402 and yes it is compatible with all leads
Is VNS open looped or close loop system?
VNS is Open looped
Competitor RNS is close looped
Describe different lead types with part numbers
302 is outside the US only and is single pin
303 is single pin and is 17 times more durable than 302
304 is the Perennia Flex 3.5 times more durable than 302
300 is compatible with the dual pin systems 102R & 104