VNS Product information Flashcards

0
Q

What parameters or 102 lead test performed at?

A
1.0 mA 
20 Hz
500 µs
30 seconds on
60 seconds off
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
1
Q

What does IFI stand for?

And what does yellow mean on different generators?

A

Intensified follow-up indicator

8 to 18% battery on 103 and 104 generators

5 to 11% battery on 105 generators

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

What are the parameters of the system diagnostic test?

A

1.0 mA
20 Hz
500 µs and
30 seconds on

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

What does N EOS mean? And by generator type?

A

Near end of service
0 to 8% battery on the 103 and 104 generators
0 to 5% battery on 105 generators

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

What are the 8 setting parameters?

A
Output current in mA
Signal frequency in Hz
Pulse width in microseconds
Signal on time 
signal off time
MAGNET:
Output current in mA
Signal frequency in Hz
Pulse width in microseconds
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

What are clinical concerns of potential end of service?

A

Increase and seizure activity
Painful or Erratic stimulation
Report of decreased perception of stimulation or no stim when using magnet

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

What are MRI conditions that are safe?

A

Implant set to 0.0 mA
Transmit and receive only
Coil type is 1.5 to 3.0 Tesla in a static magnetic field

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Describe testing impedances on 102 batteries

A

Make sure patient can tolerate 1.0 mA

System diagnostics test assesses lead impedance and integrity

Normal mode assess his generator to deliver programmed output

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

How can you tell high impedances on pulse generators? 102 or older

A

DCDC is 4-7 which coincides with high impedance

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Accessory pack part number and contents

A

Tiedowns
hex screwdriver and
2 x test resistors - one single pin and one dual pin

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

What color is sterilization indicator went good? And when bad?

A

Good = gold to bronze

Bad = Red or black

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

What is the part number of the programming wand?

A

201???? Need to verify this

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Is the tunneler compatible with all lead types? And what is the part number?

A

402 and yes it is compatible with all leads

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Is VNS open looped or close loop system?

A

VNS is Open looped

Competitor RNS is close looped

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

Describe different lead types with part numbers

A

302 is outside the US only and is single pin

303 is single pin and is 17 times more durable than 302

304 is the Perennia Flex 3.5 times more durable than 302

300 is compatible with the dual pin systems 102R & 104

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

List models of generators

A
102 is called Pulse
102R is also called Pulse but dual pin
103   Demipulse
104 Demipulse dual pin
105 - aspire high capacity
16
Q

Benefit of magnet

A

Gives control to patients and families

On-demand stim

  • avoid seizures for patients with auras
  • abort or decrease the severity
  • improve postictal period

Stop stimulation

Check if generator is still working

Titration helps patient get used to higher stimulation