Questions Flashcards
In the model 103 diagnostic test should be performed when we’re at what visits?
The first visit
List the four diagnostic tests and the purpose of each test.
Normal mode test – tests device with normal settings
Generator diagnostics – the test in the OR
Magnet mode test – test generator with magnet settings
Systems diagnostics – tests lead impedance and system
In the model 1021 reading indicates a potential problem on a system diagnostic?
7 is a high limit
In the model 102 what reading indicates a potential problem on a normal mode diagnostic?
DCDC =7
Lead impedance is high
Which diagnostic test is performed in the operating room and in the clinical setting?
System Diagnostics – OR the other three are in the clinical setting
What is the difference between a normal mode and system diagnostic?
System diagnostic assesses lead impedance and system integrity
Normal mode test the device at the parameters it is set to
In the model 105 aspire generator explain the enhanced diagnostic features:
Lead check every 24 hours
Provides ohms
Battery indicator
Won’t reset programming of generator during diagnostic
What are the normal impedance range for system diagnostics in the Demipulse generators:
600–5300 ohms
What me potentially happen to the output currently have a system diagnostic gets interrupted?
It will turn the output current to 1.0 mA
What are the parameters diagnostics are run at for the
Model 102
model 103
model 105
Model 102 - 1.0 mA
model 103 – .25 mA
model 105 – .25 mA