Ventilators and Airway monitors Flashcards
Respirometer
- In expiratory limb
- exhaled TV sensor
- activated automatically once breaths are sensed and always active during mechanical ventilation
- gas flow converted to electrical pulses
- apnea- if sufficient breath (based on TV setting) not achieved within 30 seconds
- low minute volume
Expiration phase- bellows ventilator
- drive gas exits the bellows chamber via ventilator pop off valve and pressure in chamber drops to zero
- exhaled pt gas fills the bellows before any scavenging occurs
- b/c valve ball produces a 2-3 cm H2O back pressure.
- relief (pop off) valve is ONLY open during expiration when all scavenging takes place
PSV
- Pressure support ventilation
- aid in normal breathing with predetermined level of positive pressure
- pt spontaneously breathing
- senses pt insp effort and gives pressure support
- results in larger TV than pt would produce independently
- useful to support MV and control arterial CO2 for spontaneously breathing patients during maintenance or emergence
Pressure controlled
- cuts off inspiration when pressure meets the predetermined level. TV and inspiratory time vary
possible issues with bellows
- leaks, improper seating
- hole in bellows
- hyperinflation of the lungs
- O2 concentration can change
- ventilator relief valve problems
- hypoventilation- gas goes to scavenger rather than drive
- caused by- disconnection, ruptured valve, or other damage
- valve stuck in closed position- additional peep and excess pressure
- excess suction from scavenge can also cause the valve to remain closed.
Ventilator settings
- TV: 5-6 ml/kg (ideal weight)
- RR: 8-12
- Flow rate: about 4-6 x minute ventilation
- I:E ratio: physiologic is 1:2
- Time Inspired= TV/ Flow rate
drive gas
- either air or oxygen
- using oxygen depletes supply quickly because so much goes to scavenge.
- some machines can entrain air, reducing the need for oxygen
- used in austere conditions (military)
Ventilation VOLUME
- measure of the tidal volume delivered by ventilator to patient
- volume of gas pt breaths
- expressed in mls
- Liters for minute volume
ICU vents Vs anesthesia vents
- ICU
- more powerful- greather insp pressures and tidal volumes
- more modes of ventilation
- gas supplied by the ICU ventilator directly ventilates the patient
- Anesthesia
- CO2 absorber
- driving gas never reaches the patient
- 100% O2 in older machines
- air/100% O2 in newer models
To Select Ventilation mode:
- Push the Ventilator key
- Review the settings for the new mode.
- Select Mode
- Select desired mode (mode is changed immediately upon selection)
*Ventilator is electonically controlled and pneumatically driven.
possible issues with pistons
- refill even if a circuit disconnection occurs
- if a circuit leak is present, piston ventilators may entrain RA through the leak, diluting O2 and anesthetic
- risk of hypoxemia and awareness
- an alarm will sound
- a positive-pressure relief valve on the ventilator prevents excessively high breathing circuit pressure (60-80 cm H2O)
TI equation
TI = TV/ Flow rate
feedback mechanisms for Pneumatic bellows ventilators
- help administer more stable tidal volume
- circuit compliance compensation and measure tidal volume as a feedback signal.
MV
MV = TV x RR
monitors
- ETCO2- capnography- best for revealing disconnect.
- Oxygen analyzer- most important monitor on the machine. Calibrate at 21%
- Respirometer- measures exhaled volumes
- PAP monitors- peak airway pressure
- vigilence is the best monitor
For each mmHg of PO2 …
…there is .0031 ml O2/100ml of blood
*normal arterial blood with a PO2 of 100 mmhg contains 0.3 ml of O2/100 ml
Maquet FLOW-i Anesthesia System With Volume Reflector
- uses “volume reflector”
- volume reflector is functional and “in circuit” during all modes
- expired gas gets “balanced” by Reflector gas
TE equation
TE = total time for each breath - TI
Total time for each breath:
60 sec/ RR (12 bpm) = 5 sec/breath