Scavenging system, Capnography, and CO2 absorption Flashcards
Passive gas disposal
- Pressure is raised above atmospheric by the patient exhaling, manually squeezing of the reservoir bag or ventilator
- pushes out the gas
- Waste gases are directed out of the building through an open window, pipe, extractor fan vented to the outside air
- Advantages: inexpensive, simple to operate
- disadvantages: may be impractical in some buildings
NIOSH acceptable levels of Volatile halogenated anesthetic alone
2 ppm
- Phase 1: Inspiratory baseline, no CO2. Inspiratory and first part of expiration (deadspace gas)
- Phase 2: Expiratory upstroke (mixture of dead space gas and alveolar gas). Slope determined by evenness of alveolar emptyness.
-
Phase 3: All alveolar gas. constant or slight upstroke, reading taken from highest point at end
- normal value 30-40
- Phase 4: Beginning of inspiration, rapid decline of CO2
main places where gas needing to be scavenged comes from
- APL valve***
- pressure relief valve from the ventilator ***
- gas analyzers (50-250 ml)
- gas that drives the bellows (not always, sometimes goes out into room but that is ok because it is pure oxygen)
*** = most common places scavenged from
Channeling
- exhaled gas taking same route through absorber everytime because it is path of least resistance
- results from loosely packed granules
- air space occupies 48-55% of the volume of the canister
- the absorbant along the channels will exhaust before the rest of container and it is difficult to know because they are on the inside
- use CO2 monitoring to help determine exhaustion
Lithium hydroxide reaction
2LiOH * H2O + CO2 →Li2CO3 + 3H2O - HEAT
Gas disposal tubing
- Connects the scavenging interface to the disposal assembly
- should be different in size and color from the breathing system
- with a passive system the hose should be short and wide
- Tubing running overhead ideal to prevent accidental obstruction and kinking
Calcium hydroxide lime ingredients
AKA Amsorb Plus
- 80% Calcium hydroxide
- 16% water
- 1-4% calcium chloride
- Calcium sulfate and polyvinylpyrrolidine added for hardness
*No activators for quick reaction, only the slow part
esophageal intubation
*may have some CO2 in stomach from insufflation, but it quickly declines.
NIOSH acceptable levels of Nitrous oxide
25 ppm
Baralyme info
- granules are 4-8 mesh
- slightly less efficient than soda lime but less likely to dry out.
Gas disposal assembly
components to remove gases from the OR
Two types: Active and passive
Closed Interface
Two types: Positive pressure relief only
and
Positive-pressure and negative-pressure relief
Open interface
- NO valves- open to atmosphere via holes on sides, avoids buildup of + or - pressure
- REQUIRES use of central vacuum systme and a reservoir
- Vacuum control valve can be adjusted. Level of suction must be greater than excess gas flow rate to prevent pollution
Scavenging definition
- Collection of excess gases from equipment used in administering anesthesia or exhaled by patients.
- Removal of these excess gases to an appropriate place of discharge outside the working environment.
Method of measuring CO2:
Infrared Absorption Spectrophotometry
- Most common
- Gas mixture is analyzed and propotion of its contents is determined
- each gas mixture absorbs infrared radiation at different wavelengths
- amount of CO2 is measured by detecting its absorbance at specific wavelengths and filtering the absorbance related to the other gases
scavenging interface
AKA balancing valve or balancing device
- Prevents pressure increases or decreases in the scavenging system from being transmitted to the breathing system
- interface limits pressures immediately downstream of the gas-collecting assembly to between -o.5 to +3.5 cm H20
- inlet should be 30 mm male connector
- should be as close to gas-collecting assembly as possible
PaCO2 vs ETCO2
- ETCO2 can be used to estimate PaCO2
- PaCO2 > ETCO2
- ETCO2 usually 2-3 mmhg under PaCO2
- used as an evaluation of dead space
3 basic elements of Scavenging interface
- positive pressure relief-protects patient and equipment in case of occlusion of system
- negative pressure relief limits subatmospheric pressure
- reservoir capacity matches the intermittent gas flow from gas collecting assembly to the continuous flow of disposal system.
size of absorbent granules
- 4-8 mesh
- irregular shape for increased surface area
- range in size because all too small would increase resistance but still want greatest surface area
Rebreathing capnography
*does not go back down to zero during inhalation
- Caused by:
- equipment dead space
- exhausted CO2 absorber
- inadequate fresh gas flows (in semi-open system)
Exhalation flow
- gas flows through the mask,
- into the rebrathing bag, and out the APL valve.
- (simultaneously) fresh gas continues to flow from the common gas outlet of the machine into the common gas inlet of the absorber.
- There is some retrograde flow of the fresh gas down the “return tube” and travels to the bottom of the CO absorber.
**note: exhaled gas is NOT traveling through the CO2 absorber