Wall gas hose to machine has a pressure of — to — psig for oxygen and nitrous oxide.
40-50
What is the function of the oxygen flow from wall supply check valve?
Prevents oxygen cylinder gas from escaping the anesthesia machine and going out the hospital wall. If this valve sticks open, there is no problem with hospital supplied gas but if it sticks open when on “E” cylinder gas, then “E” cylinder gas will exit out the machine via the DISS valve.
The — assembly has pins for E cylinders to fit on. The oxygen pin index is 2,5 and nitrous oxide is 3,5
Hanger yoke valve
The hanger yoke valve acts as ‘safety’ valve; — with pressure and — with pressure (free floating valve)
Opens ; closes
If the hanger yoke valve fails to open cylinder valve then the result is — gas flow to anesthesia machine.
No
What is the name of the gauge that shows oxygen tank pressure?
Bourdon
You have a dual hanger yoke with two oxygen “E” cylinders. Both cylinders are turned on; which one will suppply the gas machine?
The “E” cylinder with the highest pressure.
The Bourdon gauge shows a pressure of 825 psig; what does this mean?
There is 825 psig (pressure) from the oxygen “E” cylinder.
In a dual hanger yoke system, what prevents one cylinder from emptying into the other?
There is a free-floating valve for each cylinder and the oxygen “E” cylinder with the greatest pressure closes the free-floating valve on the cylinder with less pressure. This would hold true if one hanger yoke was empty with no cylinder.
Which oxygen cylinder pressure regulator?
Cylinder pressure decreased to 40-50 psig (intermediate pressure)
First stage regulator
Which oxygen cylinder pressure regulator?
-Reduces pressure of 40 50 psig to 16 psig (intermediate to low pressure)
-Receives wall pressure oxygen or oxygen from first stage regulator
-Diaphragm valves only reduce pressure
Second stage regulator
Oxygen flush valve (anesthetist’s panic button)
-Oxygen delivered direcly to patient at a rate of
— to — L/minute
-Pressurized at — to — psig from wall or cylinder
-Dilutes anesthetic gas - intermediate pressure
-Ball-and-spring valve - you supply the energy
-May cause barotrauma
35-75 L/minute ; 40-50 psig
Pressure sensor shut-off valve or oxygen failure pressure device
-senses oxygen pressure at — psig
-shut off — if oxygen pressure falls
-oxygen pressure must maintain at least — psig to keep valve open
-ball-and-spring type valve
-senses — only not flow delivery
50 ; nitrous oxide ; 25 ; pressure
Oxygen flow meter and flow control valve
-deliver gas to —
-receives low pressure (— psig)
-needle valve
-the — tube is gas specific and tapered with largest diameter at the top
Patient circuit ; 16 ; Thorpe
With vaporizer on, this valve is pushed — by the pressure in the line. Should a back-pressure develop that is greater than vaporizer flow pressure, the valve will — and stop the vaporizer flow.
open ; close
Check valve between vaporizer outlet and machine outlet (not present on all machines)
-Prevents — of flow to vaporizer (back pressure)
- —floating type valve (stiffer)
reversal ; Free
What happens if the inspiratory valve sticks open?
The expiratory volume will exhaust through the inspiratory limb. The ETc02 waveform will become elevated.
What happens if the expiratory valve sticks open?
Due to the path of least resistance, the inspired volume will not enter the endotracheal tube but instead will bypass and exhaust through the expiratory limb.
APL valve (pop off)
-Adjusts the limit of — in the patient circuit and rebreathing bag
-Tightening the knob will — the pressure
-Attached to the — check valve
pressure ; increase ; exhalation
Waste gas scavenging system:
-Receives gas from the — and —
-—pressure relief valve will allow gas to escape from the system to the operating room if pressure
builds up in the system
-Wall evacuation system requires the — valve to be properly adjusted. If excess flow is present, then the negative pressure relief valve will open and if flow is not enough the positive pressure relief valve opens and waste gas enters the operating room
machine (APL) and ventilator ; Positive ; needle
Adjusting the needle valve alters the flow of waste gases into the vacuum source; adjusting the needle
valve does — regulate vacuum or suction.
not
What will happen when the vaporizer is tipped then leveled?
Liquid vapor will get into the vaporizer chamber and the carrier flow will carry more agent to the patient.
What does this mean that vaporizer will carry more agent after it is tipped then leveled?
More agent picked up means a higher concentration of agent delivered to the patient. It may be a lethal dose.
What do I do if the vaporizer has been tipped then leveled?
Do NOT use the vaporizer; send it back to the company for recalibration.
The — pressure (cylinder pressure) side includes the cylinder pressure gauge (Bourdon), the check valves in the hanger yoke, and the index pins on the hanger yoke.
high
High pressure is — pressure.
cylinder
The — pressure portion of the machine includes the flush valve, second stage regulator, oxygen pressure shut-off valve, pressure gauges from pipeline, and diameter index safety system and flow control valve of manifold.
intermediate
The intermediate pressure portion can be hospital piped-in gas (— psi) or cylinder gas — the first stage regulator.
55 ; after