UT LEVEL III Flashcards
QA ASNT
- In an ultrasonic test system where signal amplitudes are displayed, an advantage of a frequency independent attenuator over a continuously variable gain control is that the:
a. pulse shape distortion is less.
b. signal amplitude measured using the attenuator
is independent of frequency.
c. dynamic range of the system is decreased.
d. effect of amplification threshold is avoided. A86
b. signal amplitude measured using the attenuator
2.An amplifier in which received echo pulses must exceed a certain threshold voltage before they can be indicated might be used to:
a. suppress amplifier noise, unimportant scatter echoes, or small discontinuity echoes that are ono consequence.
b. provide a display with nearly ideal vertical linearity characteristics.
c. compensate for the unavoidable effects of materattenuation losses.
d. provide distance-amplitude correction automatically.
a. suppress amplifier noise, unimportant scatter echoes, or small discontinuity echoes that are ono consequence.
- The output voltage from a saturated amplifier is:
a. 180° out of phase from the input voltage.
b. lower than the input voltage.
c. nonlinear with respect to the input voltage.
d. below saturation.
G.176, 182
c. nonlinear with respect to the input voltage.
- The transmitted pulse at the output of the pulser usually has a voltage of 100 to 1000 V, whereas the
voltages of the echoes at the input of the amplifier are on the order of:
a. 0.001-1 V
b. 1-5 V
c. 10 V
d. 50 V
G.174-176
a. 0.001-1 V
- The intended purpose of the adjustable calibrated attenuator of an ultrasonic instrument is to:
a. control transducer damping.
b. increase the dynamic range of the instrument.
c. broaden the frequency range.
d. attenuate the voltage applied to the transducer.
A.86
b. increase the dynamic range of the instrument.
- Which of the following might result in increased transmission of ultrasound within a coarse-grained
material?
a. Perform the examination with a smaller diameter transducer.
b. Perform the examination after a grain-refining heat treatment.
c. Change from a contact examination to an immersion examination.
d. Change from a longitudinal to a transverse wave.
B.129
b. Perform the examination after a grain-refining heat treatment.
- The term that is used to determine the relative transmittance and reflectance of ultrasonic energy at
an interface is called:
a. acoustic attenuation.
b. interface refraction.
c. acoustic impedance ratio.
d. acoustic frequency.
B.16
c. acoustic impedance ratio.
- In a forging, discontinuities associated with nonmetallic inclusions can most accurately be described as being:
a. oriented parallel to the major axis.
b. parallel to the minor axis.
c. aligned with forging flow lines.
d. oriented at approximately 45° to the forging direction.
G.340
c. aligned with forging flow lines.
- The preferred method of ultrasonically inspecting a complex-shaped forging:
a. is an automated immersion test of the finished forging using an instrument containing a calibrated attenuator in conjunction with a C-scan recorder.
b. combines thorough inspection of the billet prior to forging with a careful inspection of the finished part in all areas where the shape permits.
c. is a manual contact test of the finished part.
d. is an automated immersion test of the billet prior to forging.
F.504
b. combines thorough inspection of the billet prior to forging with a careful inspection of the finished part in all areas where the shape permits.
- When maximum sensitivity is required from a transducer:
a. a straight beam unit should be used.
b. large-diameter crystals are required.
c. the piezoelectric element should be driven at its fundamental resonant frequency.
d. the bandwidth of the transducer should be as large as possible.
A.61-62
c. the piezoelectric element should be driven at its fundamental resonant frequency.
- The sensitivity of an ultrasonic test system:
a. depends on the transducer, pulser, and amplifier used.
b. decreases as the frequency is increased.
c. increases as the resolution increases.
d. is not related to mechanical damping or the transducer.
B.46-47, 56
a. depends on the transducer, pulser, and amplifier used.
- The ability of a test system to separate the back surface echo and the echo from a small discontinuity just above this back surface:
a. depends primarily upon the pulse length generated from the instrument.
b. is not related to the surface roughness of the part under inspection.
c. is primarily related to the thickness of the part under inspection.
d. is usually improved by using a larger diameter transducer.
A.183
a. depends primarily upon the pulse length generated from the instrument.
- Transducer sensitivity is most often determined by:
a. calculations based on frequency and thickness of the piezoelectric element.
b. the amplitude of the response from an artificial discontinuity.
c. comparing it to a similar transducer made by the same manufacturer.
d. determining the ringing time of the transducer.
B.102-104, 106
b. the amplitude of the response from an artificial discontinuity.
- Side-drilled holes are frequently used as reference reflectors for:
a. distance-amplitude calibration for shear waves.
b. area-amplitude calibration.
c. thickness calibration of plate.
d. determining near-surface solutions.
A.194-198
a. distance-amplitude calibration for shear waves.
- Notches provide good reference discontinuities when a UT examination is conducted to primarily detect discontinuities such as:
a. porosity in rolled plate.
b. inadequate penetration at the root of a weld.
c. weld porosity.
d. internal inclusions.
A.197-198
b. inadequate penetration at the root of a weld.
- The difference between a compression and shear wave is:
a. quantitative measure.
b. relative particle vibration direction.
c. qualitative measure.
d. amplitude.
B.10-12; G.7
b. relative particle vibration direction.
- The particle motion for rayleigh waves is usually described as:
a. sinusoidal.
b. circular.
c. elliptical.
d. shear.
A.45; B.66; C.1-2
c. elliptical.
- Based upon wave theory and ignoring attenuation losses, the echo amplitude of a finite reflector is:
a. directly proportional to the distance to the reflector.
b. inversely proportional to the distance to the reflector.
c. directly proportional to the square of the diameter of the circular reflector.
d. inversely proportional to the square of the diameter of the circular reflector.
G.97
c. directly proportional to the square of the diameter of the circular reflector.
- The rate generator in B-scan equipment will invariably be directly connected to the:
a. display intensity circuit.
b. pulser circuit.
c. RF amplifier circuit.
d. horizontal sweep circuit.
B.3, 79-82
b. pulser circuit.
- In A-scan equipment, the RF pulser output voltage is normally in the range of:
a. 1-10 V.
b. 10-100 V.
c. 100-1000 V.
d. 1000-3000 V.
B.79
c. 100-1000 V.