Ultrasound physics Flashcards
which transducer to use
highest frequency
temporal resolution
frame rate
improved axial resolution
higher frequencies
along the axis of US beam
90 degrees to skin
Y-axis
depth resolution
improved lateral resolution
adjust focal zone
x axis - right left aspect of image
better with higher frequencies
improve temporal resolution
minimize depth
adjust frame rate
use narrow sector
benefits of compound imaging
reduces speckle
increases frame rate due to longer scan time
effect of harmonic imaging
decreases frame rate
acoustic power units
watts
watts =
joules/sec
intensity (thermal index/mechanical index)
power/beam area MW/cm2
mechanical index
likelihood of a non thermal bioeffect such as cavitation (compression and decompression of tissue)
17 mW/cm2 ophthalmic related to intensity of pulse max
value of peak negative pressure/sq root of acoustic center frequency
thermal index
ratio of the emitted acoustic power to the power to raise the temperature by 1 degree
acoustic power at depth of interest/ estimated power to elevate temp 1 degree
larger TI means larger heating potential
application specific exposure limits in mW/cm2 peripheral vascular cardiac fetal ophthalmic
720 mW/cm2 peripheral vascular
430 mW/cm2 cardiac
94 mW/cm2 fetal
17 mW/cm2 ophthalmic
according to J of US
0.23 W/cm2 (versus 1.9 for other organs)
ISPPA = spatial-peak pulse-average intensity
mechanical index
specular reflection
at smooth surfaces such as bone, diaphragm, tendons calcifications, valves, needles
angle dependent
mirror image artifact
90 degrees duplication at site of a strong reflector
far image is false image
1 reflector send it back to the other delay in sending back leads to duplicate
back scattering
associated with most tissues
less angle dependent
rough surface
Rayleigh scatter
blood - RBC’s smaller than wavelength