Tonometry CH17 Flashcards

1
Q

When cleaning the applanation tonometer, one must be careful not to:
a) remove the biprism
b) bend the pressure-sensitive arm
c) turn the adjustment knob
d) cause scleral rigidity

A

b) The moveable arm that holds the biprism can be bent if care is not taken.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

The applanation tonometer unit itself may be cleaned by:
a) immersing in 3% hydrogen peroxide
b) spraying and wiping with electronics cleaner
c) gently wiping with damp cloth and mild soap
d) spraying with disinfectant

A

c) This recommendation comes from the manufacturer.6 It is never a good idea to spray cleaner directly onto an instrument.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

The tonometer tip should be cleaned:
a) prior to being sterilized
b) prior to being autoclaved
c) prior to being boiled
d) prior to being disinfected

A

d) Technically, cleaning removes soil and debris and some germs. Disinfecting is the
removal of all or most germs except bacterial spores (which are killed by sterilization).
External debris is removed first (or at the same time) in order to allow all instrument surfaces to come into contact with the disinfectant.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

The tonometer tip should be disinfected:
a) morning and night
b) first thing each day
c) every 10 minutes
d) between each patient

A

d) Because the tonometer tip contacts the cornea and tear film, it must be disinfected (or a fresh, disinfected tip inserted) between each patient.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

To best disinfect the tonometer tip, one should:
a) wipe it with a moist tissue
b) swipe it with an alcohol wipe
c) put it in a 10-minute soak in 3% hydrogen peroxide
d) place it in the autoclave for 15 minutes

A

c) A 10-minute soak of bleach (1:10 dilution) or hydrogen peroxide is the best disinfection
method. A moist tissue will not disinfect, nor will a quick “swipe” with alcohol (although
some references suggest a 10-second wipe as acceptable). The autoclave will melt the
tonometer tip.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

With regular use, soaking the tonometer tip in alcohol:
a) will not affect it
b) will cause the numbers to fade
c) will cause etching on the face
d) will cause the plastic to soften

A

b) Repeated exposure to alcohol can cause the numbers and lines on the side of the tonometer tip to fade and disappear.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Failure to properly rinse the tonometer tip after disinfecting and prior to using it
could result in:
a) corneal chemical burn
b) inaccurate readings
c) etching of the tip’s surface
d) clouding of the mires

A

a) Alcohol and hydrogen peroxide can cause marked discomfort, so the tip should be rinsed well and dried before using.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Check calibration on the Goldmann applanation tonometer by:
a) placing the tip on a test block
b) measuring an eye of known pressure
c) use of a calibration bar
d) returning it to the manufacturer

A

c) A calibration bar is placed in a special attachable holder when calibrating an applanation tonometer. If the calibration is not accurate, the instrument must be returned to the manufacturer.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

When calibrating the Goldmann applanation tonometer at the 2 setting, the tonome-
ter head should move:
a) no more than ±0.50 from the tested position
b) 1.0 from the tested position
c) freely in every position
d) not at all

A

a) Movement of the tonometer arm should occur within 0.50 of the drum setting of 2. If
you must turn the drum more than 0.50 before the arm will move, this indicates that the instrument is not accurate. Thus, when you are calibrating at 2, the arm should “rock”
between the 1.95 and 2.05 reading on the drum.7,8

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

The test positions for calibration of the Goldmann applanation tonometer are:
a) 0, 22, and 45 mm Hg
b) 0, 20, and 60 mm Hg
c) 10, 20, and 50 mm Hg
d) 0, using a test block

A

b) The calibration bar is marked to check the 0, 20, and 60 mm Hg readings on the tonometer. Especially pay attention to the calibration at 20 mm Hg, as this is often the crucial point in patients with glaucoma.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

If you are calibrating an applanation tonometer for the 20 mm Hg setting, the drum
will read:
a) 1
b) 2
c) 4
d) 5

A

b) Remember, the drum readings are multiplied by 10, so the 20 mm Hg reading would translate to “2” on the drum.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

If the applanation tonometer falls outside of the calibration allowance:
a) add or subtract from the final reading to offset the problem
b) bend the tonometer arm until calibration is accurate
c) turn the screw in the tonometer arm until calibration is accurate
d) return the tonometer to the manufacturer for calibration

A

d) Answers a through c might sound good, but the only way to recalibrate the tonometer is to return it to the manufacturer for servicing.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

The applanation tonometer measures intraocular pressure (IOP) by:
a) measuring the amount of pressure needed to indent the cornea
b) measuring the amount of time needed to flatten the cornea
c) measuring the IOP directly
d) measuring the amount of pressure needed to flatten the cornea

A

d) The applanation tonometer flattens the cornea and measures the amount of pressure
required to do so. (By contrast, indentation tonometry measures the amount of pressure
required to indent the cornea.)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

IOP as measured by applanation is recorded as:
a) mm Hg
b) gm/mm2
c) a scale from 0 to 18
d) lb/in2

A

a) Applanation tension is measured in millimeters of mercury (mm Hg).

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

Once the cornea is applanated, the force of the applanation tonometer is increased or
decreased:
a) until the outer edges of the mires touch
b) until the edges of the mires overlap
c) until the inner edges of the mires touch
d) until the upper mire is larger

A

c) The inner edges of the mires should touch when the reading is taken.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

Wide mires in applanation tonometry will result in:
a) accurate readings
b) falsely low readings
c) falsely high readings
d) increased patient comfort

A

c) Wide mires (too much fluid) result in falsely high readings.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

Thin mires in applanation tonometry will result in:
a) accurate readings
b) falsely low readings
c) falsely high readings
d) decreased patient comfort

A

b) Thin mires (not enough fluid) result in falsely low readings.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

All of the following can cause applanation tonometry errors except:
a) too much or too little fluorescein
b) lack of contact with the eyelid
c) misalignment of mires
d) dirty tonometer face

A

b) The tonometer should not contact the eyelid at all.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

All of the following can result in inaccurate applanation readings, with no compensa-
tion method available, except:

a) astigmatism
b) pterygium
c) corneal scars
d) corneal graft

A

a) Astigmatism over 3 D must be compensated for (see answers 36-38). There is no way to set the tonometer to compensate for a pterygium, scar, or graft—all of which can cause an
inaccurate reading.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

The applanation tonometer is preferred in cases of low scleral rigidity because:
a) it does not displace an appreciable amount of aqueous and, therefore, does not cause
distention of the ocular structures
b) it does not flatten the cornea and, therefore, does not cause distention of the ocular structures
c) it is performed with the patient in a seated position and, therefore, gravity can equalize
distention of the ocular structures
d) topical anesthetic is used and, therefore, does not cause distention of the ocular structures

A

a) The applanation tonometer does not displace a significant enough amount of aqueous to cause even a more pliable eye to distend.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

The biprism design of the applanation tonometer:
a) makes it easier to align than an indentation tonometer
b) makes it more accurate than indentation tonometry
c) offsets scleral rigidity factors
d) makes it more comfortable than indentation tonometry

A

a) With the biprism, you have visual proof (the mires) of correct alignment.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Each of the following is an advantage of applanation tonometry except:
a) it gives an excellent binocular view of the mires
b) it is the most accurate method of checking IOP
c) the tonometer tip is easily cleaned and disinfected
d) it is accurate even in the presence of low scleral rigidity

A

a) The mires appear in only one ocular, making it monocular, not binocular. The other statements are true.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

Disadvantages of applanation tonometry include all of the following except:
a) it is expensive
b) slit-lamp models are not portable
c) it can be difficult to learn to use
d) one must use a chart to convert the reading

A

d) You do not need a conversion chart with applanation tonometry. The measurement is read directly off the drum.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Match the following with the drawings using each answer only once. (All drawings
reprinted with permission of Herrin MP. Ophthalmic Examination and Basic Skills.
Thorofare, NJ: SLACK Incorporated; 1990.)

proper position for measurement,
too close, reading is too low,
too far back, reading is too high
too much fluorescein,
vertical position is off,
not enough fluorescein

A

C) proper position for measurement
A) reading is too low
H) reading is too high
E) vertical position is off
D) too close
G) too far back
F) too much fluorescein
B) not enough fluorescein

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
Q

You have just realized that you instilled regular fluorescein into an eye with a soft
contact. What do you do next?
a) perform the applanation over the lens
b) remove the lens, and rinse it immediately
c) irrigate the eye with the lens in it
d) slide the lens to the side, and take the measurement

A

b) If you remove the lens and rinse it right away, you may be able to get the dye out before the lens is too stained.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
26
Q

For proper applanation, the slit-lamp light should be set up as follows:
a) at a 60 degree angle, with the cobalt blue filter, and with the light source completely
open
b) at a 45 degree angle, with the red-free filter, and with the light source completely open
c) at a 90 degree angle, with the cobalt blue filter, and a narrow beam
d) at a 60 degree angle, with the cobalt blue filter, and a pinpoint light beam

A

a) To perform applanation tonometry, the light source should be at a 60 degree angle with the cobalt blue filter and full illumination.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
27
Q

During applanation tonometry, the patient is instructed to:
a) close the eye not being tested, hold his or her breath, and look up
b) open both eyes, breathe normally, and look to the right
c) open both eyes, breathe normally, and look straight ahead
d) open both eyes, hold his or her breath, and look straight ahead

A

c) Applanation tensions are easier if the patient will open both eyes. The patient must look straight ahead for proper alignment. The patient should not hold his or her breath, as this can cause a falsely elevated IOP reading.

28
Q

Each of the following can be key in accurate applanation tonometry except:
a) the patient should loosen a tight collar
b) the patient should clamp his or her teeth
c) the patient should not strain to reach the slit lamp
d) the patient should relax the shoulders and neck

A

b) The reading will be most accurate if the patient is relaxed.

29
Q

Your patient is a 45-year-old executive. You have just taken his applanation tonometry
reading, and it is 28. There is no history of glaucoma in the family. His readings for
the past 10 years at annual exams in your office have been 20. You lean back, scratch
your head, then ask the patient to:

a) come back for a reading in the afternoon
b) drink a lot of water to see if this will affect the pressure reading
c) loosen his tie and collar, then repeat the measurement
d) let you recheck with the Schiøtz tonometer

A

c) Have him loosen his tie and collar and try again. (The physician would be the one to
order answers a or b.) The Schiøtz is less accurate than the applanation tonometer.

30
Q

If, during applanation tonometry, it is noticed that the mires are not aligned, the
proper procedure is to:
a) adjust the joystick to realign
b) adjust the pressure reading accordingly
c) adjust the pressure reading using the appropriate chart
d) back off the eye, adjust the joystick, then reapplanate

A

d) Pull the tonometer off the cornea just slightly before adjusting position. Do not slide the tip around on the cornea.

31
Q

Each mark on the applanation tonometer drum stands for:
a) 10 mm Hg
b) 2 mm Hg
c) 1 mm Hg
d) 2 g/mm2

A

b) Each mark on the tonometer drum represents 2 mm Hg.

31
Q

Between measuring the two eyes, it is helpful to ask the patient to:
a) look left then right to help him or her relax
b) blink to redistribute the fluorescein
c) blot his or her eyes with a tissue
d) look straight ahead without blinking

A

b) We ask the patient not to blink during the test, so the second eye may tend to be a little
dry. A quick blink will spread the dye across the cornea.

32
Q

The reading on the applanation tonometer drum is 2. The patient’s IOP is:
a) 20 mm Hg
b) 4 mm Hg
c) 2 mm Hg
d) 4 g/mm2

A

a) If the number 2 on the drum is on the marker line, the IOP is 20. Each mark is 2 mm Hg, but each number is 10 mm Hg.

33
Q

In properly aligned applanation tonometry:

a) there are no factors that affect the IOP
b) the mires will pulsate slightly with the heartbeat
c) the mires will pulsate slightly when the patient breathes
d) accuracy is ensured if the patient does not blink

A

b) The mires may pulsate slightly with the heartbeat (not with breathing). Even with accurate alignment, however, there are other factors that can induce error.

34
Q

When the applanation tonometer mires pulsate, the IOP:
a) is the lowest of the readings
b) is the highest of the readings
c) is the difference between the 2 readings
d) is at the midpoint between the 2 readings

A

d) Take the reading at midpoint between the readings during the pulsations.

35
Q

One must make an adjustment to the applanation tonoprism if the patient’s corneal
astigmatism is:
a) greater than 1 D
b) greater than 2 D
c) greater than 3 D
d) greater than 4 D

A

c) The “magic number” is 3 D when it comes to resetting the tonoprism for astigmatism.
Failure to do so may result in an erroneous measurement (see answers 37 and 38 for more information).

36
Q

If compensation for high corneal astigmatism is not made to the applanation tonom-
eter, the IOP measurement could be in error by:
a) 1 mm Hg
b) 2 to 3 mm Hg
c) 4 to 5 mm Hg
d) 8 to 10 mm Hg

A

b) If the biprism is not adjusted properly for the astigmatic cornea (more than 3 D), the IOP measurement could be 2 to 3 mm Hg off.

37
Q

In order to compensate for high astigmatism with the applanation tonometer, the
biprism should be aligned as follows:
a) the steepest axis aligned with the red line on the holder
b) the plus axis aligned with the red line on the holder
c) 45 degrees from the minus cylinder should be placed in the 90-degree position
d) the minus axis aligned with the red line on the holder

A

d) Turn the biprism so that the axis (in minus cylinder) is in line with the red mark on the
tonometer holder. It is best to get the cylinder reading from the keratometer rather than the
patient’s glasses prescription, because the glasses may also correct for lenticular astigmatism. (Note: Some sources say to turn the biprism 45 degrees from the flattest axis, which is the point indicated by the red line on the holder.)

38
Q

A scarred, irregular cornea is difficult to measure with the applanation tonometer
because:
a) there is decreased scleral rigidity
b) one cannot instill topical anesthetic because of tissue melt
c) one cannot use fluorescein because it will infiltrate the tissue
d) the mires are irregular, making it difficult to judge the endpoint

A

d) If the cornea is irregular, the mires will also be irregular. This makes it tough to tell when
the inner edges are meeting.

39
Q

The most common complication of contact tonometry is:
a) fainting
b) allergy to anesthetic drops
c) cardiac arrest
d) conjunctivitis

A

d) Spreading conjunctivitis from one patient to the next is the most common complication
of contact tonometry.

40
Q

Corneal abrasions from tonometry can best be reduced by:
a) use of anesthetic
b) proper fixation
c) proper lighting
d) holding the lids securely

A

b) If the patient will fixate properly, holding the eye still, the risk of corneal abrasion is
minimized. (Holding the lids securely would help, but b is the best answer.)

41
Q

Measuring IOP is contraindicated in any patient:
a) who is nervous about the procedure
b) who has possible optic nerve disease
c) who has vision below 20/200
d) who might have a penetrating injury

A

d) A patient with a penetrating injury has an opening for bacteria to be introduced into the
eye via fingers, eye drops, tonometer tip, etc. In addition, aqueous could be forced out of
the eye during applanation.

42
Q

Any type of contact tonometry must be avoided, if possible, in which patient?
a) patient with acquired immune deficiency syndrome (AIDS)
b) patient with a corneal graft
c) patient with epidemic keratoconjunctivitis
d) children younger than 8 years of age

A

c) Epidemic keratoconjunctivitis is extremely contagious! There is no contraindication in
any of the other cases.

43
Q

If the patient has known human immunodeficiency virus (HIV), the assistant should:
a) refuse to take a pressure on the patient
b) use only the noncontact tonometer
c) wear gloves during contact tonometry
d) place a tissue on the tonometer prism

A

c) HIV has been isolated from human tears, so gloving is a wise precaution. (However, to
my knowledge, there have been no reports of the virus being spread through contact with
tears.)

44
Q

A patient who reports an allergy to topical anesthetics:
a) should have his or her IOP checked with a noncontact tonometer
b) should be irrigated after contact tonometry if topical anesthetic is used
c) cannot have a pressure check of any kind
d) should tough out contact tonometry without drops

A

a) Use the noncontact tonometer on the patient who is sensitive to topical anesthetic, and you cannot go wrong. Answer b could be done if your doctor orders it. (The words “can-
not…of any kind” should have clued you in that answer c was wrong.) I have done an
occasional applanation tension using tear drops only, but it is not ideal.

45
Q

All of the following can result in a falsely high IOP reading except:
a) breath holding
b) moving the eye during the measurement
c) straining to lean into the slit lamp
d) squeezing the eyelids

A

b) Moving the eye during measurement changes alignment but does not falsely elevate the reading.

46
Q

Schiøtz tonometry measures IOP by:
a) measuring the force it takes to equalize the eye’s internal pressure
b) measuring the force exerted on the instrument by the eye
c) measuring the force it takes to flatten an area of the cornea
d) measuring the force it takes to indent the cornea

A

d) Schiøtz tonometry measures IOP by indentation.

47
Q

The Schiøtz tonometer readings:
a) are read from the instrument in millimeters of mercury
b) are read from the instrument in gram weights
c) must be converted to millimeters of fluorescein by use of a calculator
d) must be converted to millimeters of mercury by use of a chart

A

d) The Schiøtz reading must be converted to mm Hg by use of a graph or chart.

48
Q

The Schiøtz tonometer should be disinfected:
a) at the beginning of each day
b) at the end of each day
c) between each patient
d) when it appears dirty

A

c) Because the Schiøtz comes in direct contact with the cornea and tear film, it must be
disinfected between each patient.

49
Q

Patient position for Schiøtz tonometry is:
a) sitting up to the slit lamp
b) sitting up in the exam chair
c) lying back in the exam chair
d) sitting at a 45 degree angle

A

c) The patient must lie flat on his or her back for the reading.

50
Q

The patient is apprehensive about having that “air puff” test. She is most likely refer-
ring to:
a) the TonoPen
b) the corneal sensitivity test
c) a corneal biopsy
d) noncontact tonometry

A

d) Noncontact tonometry is often called the “air puff” test.

51
Q

The noncontact, or “air puff,” tonometer measures:
a) the IOP directly
b) either the time or force required to indent the cornea
c) either the time or force required to flatten the cornea
d) either the time or force required to photograph the cornea

A

c) The noncontact tonometer may measure either time or force required to flatten the cornea.

52
Q

The noncontact tonometer is most useful for:
a) monitoring patients with glaucoma
b) differentiating between ocular hypertension and glaucoma
c) finding patients with low-tension glaucoma
d) screening situations

A

d) The noncontact tonometer is adequate for screening, but will miss some elevated pres-
sures. Thus, it is not sensitive enough for monitoring or diagnosing glaucoma.

53
Q

The aqueous humor is produced by the:
a) ciliary muscle
b) ciliary body
c) trabecular meshwork
d) iris

A

b) The ciliary body, at the base of the iris (but not the iris itself), is responsible for aqueous
production.

54
Q

The composition of the aqueous is most like:
a) blood plasma
b) tears
c) saliva
d) mucus

A

a) Blood plasma is most like aqueous in composition, with some difference in trace elements.

55
Q

The flow of aqueous in the eye follows this pattern:
a) angle, posterior chamber, pupil, anterior chamber
b) angle, anterior chamber, pupil, posterior chamber
c) pupil, posterior chamber, anterior chamber, angle
d) posterior chamber, pupil, anterior chamber, angle

A

d) The aqueous flows out of the posterior chamber, through the pupil, into the anterior
chamber, and then through the angle.

56
Q

As it exits the eye, aqueous humor flows in this pattern:
a) canal of Schlemm, trabecular meshwork, episcleral arteries
b) trabecular meshwork, canal of Schlemm, episcleral veins
c) trabecular meshwork, nasolacrimal duct, episcleral arteries
d) canal of Schlemm, episcleral veins, trabecular meshwork

A

b) The flow of aqueous humor exiting the angle goes through the trabecular meshwork, into the canal of Schlemm, and from there into the episcleral veins. (Remember that veins carry blood to the heart. Arteries bring blood to an organ.)

57
Q

IOP is determined by:
a) systolic and diastolic blood pressure
b) rate of aqueous production and resistance to outflow
c) pressure in the ophthalmic artery and vein
d) cranial pressure transferred to the eye through the optic nerve

A

b) IOP results from the combination of the amount of aqueous produced over time and its ability to drain out of the eye.

58
Q

Reduction and control of elevated IOP is based on:
a) lowering the diastolic and systolic blood pressure
b) lowering cranial pressure
c) increasing aqueous production and/or decreasing outflow
d) decreasing aqueous production and/or increasing outflow

A

d) In order to lower the IOP (theoretically, at least), decrease the amount of aqueous being
produced and/or increase the draining of it.

59
Q

Which of the following regarding aqueous and IOP is not true?
a) IOP is generally higher in the morning than in the evening.
b) IOP is slightly higher in the posterior chamber than in the anterior chamber.
c) Aqueous has no effect on the optical system of the eye.
d) Aqueous provides nutrition and waste removal for internal ocular structures.

A

c) While slight, the aqueous does exert a refractive influence on light entering the eye. The IOP is slightly higher in the posterior chamber because the aqueous meets some resistance as it encounters the margin of the pupil. The circulating aqueous also acts to bring nutrients to and remove wastes from the eye’s internal structures. Regarding answer a, see answer 61.

ANSWER 61: IOP varies during the day (“diurnal”). It is usually higher early in the morning and lower during the afternoon. (In monitoring glaucoma, many physicians will vary the time of day that a patient comes in for IOP checks, to get a clearer idea of how treatment is affecting the pressure.)

60
Q

Regarding the normal diurnal variation of IOP:
a) IOP is higher early in the morning
b) IOP is lower early in the morning
c) IOP reaches a peak around noon
d) IOP is lowest around noon

A

a) IOP varies during the day (“diurnal”). It is usually higher early in the morning and lower
during the afternoon. (In monitoring glaucoma, many physicians will vary the time of day that a patient comes in for IOP checks, to get a clearer idea of how treatment is affecting the pressure.)

61
Q

The diurnal curve of IOP in a glaucoma patient:
a) may vary by 4 mm Hg
b) may vary up to 10 mm Hg
c) is less than in an eye without glaucoma
d) shows no variation in morning versus evening

A

b) The diurnal curve of the IOP in a patient with glaucoma may vary up to 10 mm Hg, being higher in the morning. The average (nonglaucomatous) eye varies only by about 4 mm Hg. Thus, this larger fluctuation is a factor in both diagnosis and treatment of glaucoma.

62
Q

Elevated IOP as seen in chronic open-angle glaucoma is believed to be the result of:
a) decreased function of cells in the trabecular meshwork
b) obstruction of the trabecular meshwork by particles
c) overproduction of aqueous
d) optic nerve damage

A

a) The degeneration of the trabecular meshwork (in function and/or cell density) as seen in open-angle glaucoma is not a product of the normal aging process. It seems to be disease-specific. Obstruction by particulate matter would be a secondary glaucoma. While over-production of aqueous does exist, it is not the primary entity in open-angle glaucoma.

63
Q

Patients who experience an increase in IOP while using corticosteroids are called:
a) ocular hypertensives
b) glaucoma suspects
c) steroid regulators
d) steroid responders

A

d) Steroids do not cause a rise in IOP in every patient, but those patients who do experience an IOP increase while on steroids are known as steroid responders.

64
Q

If the cornea is extremely thick or scarred:
a) the IOP measurements will always be underestimated
b) the IOP measurements will always be overestimated
c) the IOP measurements can be accepted with reservations
d) no tonometric measurement will be accurate enough to satisfy clinical needs

A

d) This opinion comes from Brubaker in his excellent article entitled Tonometry.
10 Other sources agree.

65
Q

You are about to do applanation tonometry on a patient, and she asks, “Will this tell
you if I have glaucoma?” You answer:
a) “Yes, but Dr. Jackson will give you the results.”
b) “Only if the reading is abnormal.”
c) “No, this is just a screening.”
d) “Eye pressure is just one part of diagnosing glaucoma.”

A

d) Although often called “the glaucoma test,” tonometry by itself is not adequate to diagnose glaucoma.