Tonometry CH17 Flashcards
When cleaning the applanation tonometer, one must be careful not to:
a) remove the biprism
b) bend the pressure-sensitive arm
c) turn the adjustment knob
d) cause scleral rigidity
b) The moveable arm that holds the biprism can be bent if care is not taken.
The applanation tonometer unit itself may be cleaned by:
a) immersing in 3% hydrogen peroxide
b) spraying and wiping with electronics cleaner
c) gently wiping with damp cloth and mild soap
d) spraying with disinfectant
c) This recommendation comes from the manufacturer.6 It is never a good idea to spray cleaner directly onto an instrument.
The tonometer tip should be cleaned:
a) prior to being sterilized
b) prior to being autoclaved
c) prior to being boiled
d) prior to being disinfected
d) Technically, cleaning removes soil and debris and some germs. Disinfecting is the
removal of all or most germs except bacterial spores (which are killed by sterilization).
External debris is removed first (or at the same time) in order to allow all instrument surfaces to come into contact with the disinfectant.
The tonometer tip should be disinfected:
a) morning and night
b) first thing each day
c) every 10 minutes
d) between each patient
d) Because the tonometer tip contacts the cornea and tear film, it must be disinfected (or a fresh, disinfected tip inserted) between each patient.
To best disinfect the tonometer tip, one should:
a) wipe it with a moist tissue
b) swipe it with an alcohol wipe
c) put it in a 10-minute soak in 3% hydrogen peroxide
d) place it in the autoclave for 15 minutes
c) A 10-minute soak of bleach (1:10 dilution) or hydrogen peroxide is the best disinfection
method. A moist tissue will not disinfect, nor will a quick “swipe” with alcohol (although
some references suggest a 10-second wipe as acceptable). The autoclave will melt the
tonometer tip.
With regular use, soaking the tonometer tip in alcohol:
a) will not affect it
b) will cause the numbers to fade
c) will cause etching on the face
d) will cause the plastic to soften
b) Repeated exposure to alcohol can cause the numbers and lines on the side of the tonometer tip to fade and disappear.
Failure to properly rinse the tonometer tip after disinfecting and prior to using it
could result in:
a) corneal chemical burn
b) inaccurate readings
c) etching of the tip’s surface
d) clouding of the mires
a) Alcohol and hydrogen peroxide can cause marked discomfort, so the tip should be rinsed well and dried before using.
Check calibration on the Goldmann applanation tonometer by:
a) placing the tip on a test block
b) measuring an eye of known pressure
c) use of a calibration bar
d) returning it to the manufacturer
c) A calibration bar is placed in a special attachable holder when calibrating an applanation tonometer. If the calibration is not accurate, the instrument must be returned to the manufacturer.
When calibrating the Goldmann applanation tonometer at the 2 setting, the tonome-
ter head should move:
a) no more than ±0.50 from the tested position
b) 1.0 from the tested position
c) freely in every position
d) not at all
a) Movement of the tonometer arm should occur within 0.50 of the drum setting of 2. If
you must turn the drum more than 0.50 before the arm will move, this indicates that the instrument is not accurate. Thus, when you are calibrating at 2, the arm should “rock”
between the 1.95 and 2.05 reading on the drum.7,8
The test positions for calibration of the Goldmann applanation tonometer are:
a) 0, 22, and 45 mm Hg
b) 0, 20, and 60 mm Hg
c) 10, 20, and 50 mm Hg
d) 0, using a test block
b) The calibration bar is marked to check the 0, 20, and 60 mm Hg readings on the tonometer. Especially pay attention to the calibration at 20 mm Hg, as this is often the crucial point in patients with glaucoma.
If you are calibrating an applanation tonometer for the 20 mm Hg setting, the drum
will read:
a) 1
b) 2
c) 4
d) 5
b) Remember, the drum readings are multiplied by 10, so the 20 mm Hg reading would translate to “2” on the drum.
If the applanation tonometer falls outside of the calibration allowance:
a) add or subtract from the final reading to offset the problem
b) bend the tonometer arm until calibration is accurate
c) turn the screw in the tonometer arm until calibration is accurate
d) return the tonometer to the manufacturer for calibration
d) Answers a through c might sound good, but the only way to recalibrate the tonometer is to return it to the manufacturer for servicing.
The applanation tonometer measures intraocular pressure (IOP) by:
a) measuring the amount of pressure needed to indent the cornea
b) measuring the amount of time needed to flatten the cornea
c) measuring the IOP directly
d) measuring the amount of pressure needed to flatten the cornea
d) The applanation tonometer flattens the cornea and measures the amount of pressure
required to do so. (By contrast, indentation tonometry measures the amount of pressure
required to indent the cornea.)
IOP as measured by applanation is recorded as:
a) mm Hg
b) gm/mm2
c) a scale from 0 to 18
d) lb/in2
a) Applanation tension is measured in millimeters of mercury (mm Hg).
Once the cornea is applanated, the force of the applanation tonometer is increased or
decreased:
a) until the outer edges of the mires touch
b) until the edges of the mires overlap
c) until the inner edges of the mires touch
d) until the upper mire is larger
c) The inner edges of the mires should touch when the reading is taken.
Wide mires in applanation tonometry will result in:
a) accurate readings
b) falsely low readings
c) falsely high readings
d) increased patient comfort
c) Wide mires (too much fluid) result in falsely high readings.
Thin mires in applanation tonometry will result in:
a) accurate readings
b) falsely low readings
c) falsely high readings
d) decreased patient comfort
b) Thin mires (not enough fluid) result in falsely low readings.
All of the following can cause applanation tonometry errors except:
a) too much or too little fluorescein
b) lack of contact with the eyelid
c) misalignment of mires
d) dirty tonometer face
b) The tonometer should not contact the eyelid at all.
All of the following can result in inaccurate applanation readings, with no compensa-
tion method available, except:
a) astigmatism
b) pterygium
c) corneal scars
d) corneal graft
a) Astigmatism over 3 D must be compensated for (see answers 36-38). There is no way to set the tonometer to compensate for a pterygium, scar, or graft—all of which can cause an
inaccurate reading.
The applanation tonometer is preferred in cases of low scleral rigidity because:
a) it does not displace an appreciable amount of aqueous and, therefore, does not cause
distention of the ocular structures
b) it does not flatten the cornea and, therefore, does not cause distention of the ocular structures
c) it is performed with the patient in a seated position and, therefore, gravity can equalize
distention of the ocular structures
d) topical anesthetic is used and, therefore, does not cause distention of the ocular structures
a) The applanation tonometer does not displace a significant enough amount of aqueous to cause even a more pliable eye to distend.
The biprism design of the applanation tonometer:
a) makes it easier to align than an indentation tonometer
b) makes it more accurate than indentation tonometry
c) offsets scleral rigidity factors
d) makes it more comfortable than indentation tonometry
a) With the biprism, you have visual proof (the mires) of correct alignment.
Each of the following is an advantage of applanation tonometry except:
a) it gives an excellent binocular view of the mires
b) it is the most accurate method of checking IOP
c) the tonometer tip is easily cleaned and disinfected
d) it is accurate even in the presence of low scleral rigidity
a) The mires appear in only one ocular, making it monocular, not binocular. The other statements are true.
Disadvantages of applanation tonometry include all of the following except:
a) it is expensive
b) slit-lamp models are not portable
c) it can be difficult to learn to use
d) one must use a chart to convert the reading
d) You do not need a conversion chart with applanation tonometry. The measurement is read directly off the drum.
Match the following with the drawings using each answer only once. (All drawings
reprinted with permission of Herrin MP. Ophthalmic Examination and Basic Skills.
Thorofare, NJ: SLACK Incorporated; 1990.)
proper position for measurement,
too close, reading is too low,
too far back, reading is too high
too much fluorescein,
vertical position is off,
not enough fluorescein
C) proper position for measurement
A) reading is too low
H) reading is too high
E) vertical position is off
D) too close
G) too far back
F) too much fluorescein
B) not enough fluorescein
You have just realized that you instilled regular fluorescein into an eye with a soft
contact. What do you do next?
a) perform the applanation over the lens
b) remove the lens, and rinse it immediately
c) irrigate the eye with the lens in it
d) slide the lens to the side, and take the measurement
b) If you remove the lens and rinse it right away, you may be able to get the dye out before the lens is too stained.
For proper applanation, the slit-lamp light should be set up as follows:
a) at a 60 degree angle, with the cobalt blue filter, and with the light source completely
open
b) at a 45 degree angle, with the red-free filter, and with the light source completely open
c) at a 90 degree angle, with the cobalt blue filter, and a narrow beam
d) at a 60 degree angle, with the cobalt blue filter, and a pinpoint light beam
a) To perform applanation tonometry, the light source should be at a 60 degree angle with the cobalt blue filter and full illumination.