Temporal Resolution and Real-Time Imaging Flashcards
Which of the following will improve a system’s temporal resolution?
a. increased sector angle
b. increased line density
c. increased PRF
d. increased frequency
c. increased PRF
True or False?
An ultrasound system with a longer pulse duration will generally have better temporal resolution.
False
True or False?
The length of a pulse does not directly influence the temporal resolution.
True
Two imaging systems produce acoustic pulses; one pulse is 0.4 usec long and the other is 0.2 usec long. Which is most likely to provide the best temporal resolution?
a. 0.4 usec system
b. 0.2 usec system
c. they are the same
d. cannot be determined
d. cannot be determined
Which two factors determine temporal resolution?
a. speed of sound and depth of view
b. pixel size and sector size
c. focal zones and sector depth
d. propagation speed and frame rate
a. speed of sound and depth of view
Which two controls determine a system’s temporal resolution?
a. frame rate and sector depth
b. focal zones and sector depth
c. propagation speed and frame rate
d. depth of view and pulses per image
d. depth of view and pulses per image
True or False?
An ultrasound system with a shorter pulse repetition period (PRP) will have better temporal resolution than a system with a longer PRP.
True
What is the primary disadvantage of multiple focal zones along each scan line of a two-dimensional image?
a. improved temporal resolution
b. decreased temporal resolution
c. improved lateral resolution
d. improved longitudinal resolution
b. decreased temporal resolution
What is the primary advantage of multiple focal zones along each scan line of two-dimensional image?
a. improved temporal resolution
b. decreased temporal resolution
c. improved lateral resolution
d. improved longitudinal resolution
c. improved lateral resolution
Which choice will result in increased frame rate and temporal resolution?
a. decreased propagation speed
b. decreased sector size
c. decreased lateral resolution
d. decreased pulse repetition frequency
b. decreased sector size
True or False?
With sector scanning, images with greater detail (spatial resolution) are created when the number of acoustic pulses per degree of sector is increased.
True
An ultrasound system with a 4.0 MHz transducer is used to image structures as deep as 15 cm. Twenty images are produced each second, each requiring 100 acoustic pulses. What is the PRF of the system?
a. 1,500 Hz
b. 300 Hz
c. 2,000 Hz
d. 4 MHz
c. 2,000 Hz
Which of the following sector imaging systems will have the best image detail if all other parameters are identical?
a. a 90 degree sector with 100 pulses/image
b. an 80 degree sector with 40 pulses/image
c. a 70 degree sector with 84 pulses/image
d. a 60 degree sector with 60 pulses/image
c. a 70 degree sector with 84 pulses/image
Of the following, which imaging modality has the poorest temporal resolution?
a. two-dimensional, real time
b. A-mode
c. color flow imaging
d. M-mode
c. color flow imaging
Which ultrasound imaging modality has the best temporal resolution?
a. B-scanning
b. duplex imaging
c. color flow imaging
d. M-mode
d. M-mode